Operation – Teledyne 3000PB - Bulkhead mount percent oxygen analyzer User Manual

Page 41

background image

Percent Oxygen Analyzer

Operation 4

4-1

Teledyne Analytical Instruments

Operation

4.1

Introduction

Once the analyzer has been installed, it can be configured for your

application. To do this you will:

Set system parameters:

Establish a security password, if desired, requiring operator to
log in.

Establish and start an automatic calibration cycle, if desired.

Calibrate the instrument.

Define the three user selectable analysis ranges. Then choose
autoranging or select a fixed range of analysis, as required.

Set alarm setpoints, and modes of alarm operation (latching,
failsafe, etc).

Before you configure your 3000PB these default values are in effect:

Ranges: LO = 1 %, MED = 5 %, HI = 10 %
Auto Ranging: ON
Alarm Relays: Defeated, 10 %, HI, Not failsafe, Not latching
Zero: Auto, every 0 days at 0 hours
Span: Auto, at 20.9 %, every 0 days at 0 hours

If you choose not to use password protection, the default password is

automatically displayed on the password screen when you start up, and you
simply press

Enter

for access to all functions of the analyzer.