3 Installation

Model 3000PB

3-14

Teledyne Analytical Instruments

and calibration gases. These valves are completely under control of the
3000PB Electronics. They can be externally controlled only indirectly
through the Remote Cal Inputs, described below.

Pressure, flow, and safety considerations are the same as prescribed for

the SAMPLE IN inlet, above.

EXHAUST OUT: Exhaust connections must be consistent with the

hazard level of the constituent gases. Check Local, State, and Federal laws,
and ensure that the exhaust stream vents to an appropriately controlled area if
required.

Note: If your 3000PB has the –V option, see Sample In, above, for

gas pressure/flow considerations.

3.6

Testing the System

Before plugging the instrument into the power source:

Check the integrity and accuracy of the gas connections. Make
sure there are no leaks.

Check the integrity and accuracy of the electrical connections.
Make sure there are no exposed conductors

Check that sample pressure is between 3 and 40 psig, according
to the requirements of your process.

Check that the voltage selector switch on the Electrical
Connector Panel is in the appropriate position for your power
source.

Power up the system, and test it by performing the following

operations:

1. Repeat the Self-Diagnostic Test as described in chapter 4, section

4.3.5.