4 Operation

Model 3000PB

4-18

Teledyne Analytical Instruments

The analog output signal has a voltage which depends on the oxygen

concentration AND the currently activated analysis range. To relate the
signal output to the actual concentration, it is necessary to know what range
the instrument is currently on, especially when the analyzer is in the
autoranging mode.

To provide an indication of the range, a second pair of analog output

terminals are used. They generate a steady preset voltage (or current if you
have current outputs) to represent a particular range. The following table
gives the range ID output for each analysis range:

Range

Voltage (V)

Current (mA)

LO

0.25

8

MED

0.50

12

HI

0.75

16

CAL (0-25%)

1.00

20