4 Operation

Model 3000PB

4-2

Teledyne Analytical Instruments

4.2

Using the Data Entry and Function

Buttons

Data Entry Buttons: The < > arrow buttons select options from the

menu currently displayed on the VFD screen. The selected option blinks.

When the selected option includes a modifiable item, the

ΔΔΔΔΔ∇ arrow

buttons can be used to increment or decrement that modifiable item.

The

Enter

button is used to accept any new entries on the VFD screen.

The

Escape

button is used to abort any new entries on the VFD screen that

are not yet accepted by use of the

Enter

button.

Figure 4-1 shows the hierarchy of functions available to the operator via

the function buttons. The six function buttons on the analyzer are:

Analyze.

This is the normal operating mode. The analyzer

monitors the oxygen content of the sample, displays the percent
of oxygen, and warns of any alarm conditions.

System.

The system function consists of six subfunctions that

regulate the internal operations of the analyzer:

Set LCD screen contrast

Setup Auto-Cal

Assign Password

Initiate Self -Test

Check software version

View sensor output

Log out.

Zero

. Used to set up a zero calibration.

Span.

Used to set up a span calibration.

Alarms.

Used to set the alarm setpoints and determine whether

each alarm will be active or defeated, HI or LO acting, latching,
and/or failsafe.

Range.

Used to set up three analysis ranges that can be switched

automatically with auto-ranging or used as individual fixed
ranges.

Any function can be selected at any time by pressing the appropriate

button (unless password restrictions apply). The order as presented in this
manual is appropriate for an initial setup.

Each of these functions is described in greater detail in the following

procedures. The VFD screen text that accompanies each operation is repro-

Contrast Function is DISABLED

(Refer to Section 1.6)