Teledyne 3000PB - Bulkhead mount percent oxygen analyzer User Manual
Page 39

Percent Oxygen Analyzer
Installation 3
3-13
Teledyne Analytical Instruments
1. Insert the tube into the tube fitting, and finger-tighten the nut until
the tubing cannot be rotated freely, by hand, in the fitting. (This
may require an additional
1
/
8
turn beyond finger-tight.)
2. Hold the fitting body steady with a backup wrench, and with
another wrench rotate the nut another 1
1
/
4
turns.
Figure 3-10: Gas Connector Panel
SAMPLE IN: In the standard model, gas connections are made at the
SAMPLE IN
and
EXHAUST OUT
connections. Calibration gases must be
Tee'd into the Sample inlet with appropriate valves.
The gas pressure in should be reasonably regulated. Pressures between
3 and 40 psig are acceptable as long as the pressure, once established, will
keep the front panel flowmeter reading in an acceptable range (0.1 to 2.4
SLPM). Exact figures will depend on your process.
If greater flow is required for improved response time, install a bypass
in the sampling system upstream of the analyzer input.
Note: If you have the –V option, The above numbers apply instead to
the vacuum at the EXHAUST OUT connector, described be-
low, with minus signs before the pressure readings.
ZERO IN and SPAN IN: These are additional ports, included on
models with the -C option, for inputting span gas and zero gas. There are
electrically operated valves inside for automatic switching between sample
Behind Panel