Percent Oxygen Analyzer

Maintenance 5

5-5

Teledyne Analytical Instruments

2. Unlatch and open the outer and inner doors.

3. Leave the cell block installed. Pull up on the probe, with a slight

rocking motion, to release it from the probe receptacle.

4. Do not remove the O-rings unless they are worn and no longer hold the

probe tightly. (If worn, replace them.)

Figure 5-2: Exploded View of Cell Block and Micro-Fuel Cell

5. When it is free, unscrew the cap from the nylon probe. Hold the probe

vertically to prevent dropping the cell out of the probe.