5 Maintenance

Model 2000A-EU

5-2

Teledyne Analytical Instruments

Preamp

0

OK

1

Zero too high

2

Amplifier output doesn't match test input

3

Both Failed

Cell

0

OK

1

Failed (open filament, short to ground, no
power.)

2

Unbalance (deterioration of filaments, blocked
tube)

5.3

VFD Display

NOTE:

Vaccum Fluorescent Display is used. It does not need
contrast adjustment.

If you cannot read anything on the VFD, especially after first powering

up, check thead VFD cable is not loose.

5.4

Fuse Replacement

1. Place small screwdriver in notch, and pry cover off, as shown in

Figure 5-1.

Figure 5-1: Removing Fuse Block from Housing

2. To change between American and European fuses, remove the

single retaining screw, flip Fuse Block over 180 degrees, and
replace screw.