beautypg.com

Teledyne 2000A-EU - General purpose thermal conductivity analyzer User Manual

Page 50

background image

4 Operation

Model 2000A-EU

4-12

Teledyne Analytical Instruments

Note: Shut off the gas pressure before connecting it to the analyzer,

and be sure to limit pressure to 40 psig or less when turning it
back on.

Readjust the gas pressure into the analyzer until the flowrate through the

sensor settles between 50 to 200 cc/min (approximately 0.1 to 0.4 scfh).

4.4.1 Zero Cal

The

Zero

button on the front panel is used to enter the zero calibration

function. Zero calibration can be performed in either the automatic or manual
mode.

CAUTION: If you are zeroing the Cal Range by itself (multiple

application analyzers only), use

manual mode

zeroing.

If you want to calibrate ALL of the ranges at once
(multiple application analyzers only), use

auto mode

zeroing in the Cal Range.

Make sure the zero gas is flowing to the instrument. If you get a

CELL

CANNOT BE BALANCED message while zeroing skip to section 4.4.1.3.

4.4.1.1

Auto Mode Zeroing

Observe the precautions in sections 4.4 and 4.4.1, above. Press

Zero

to

enter the zero function mode. The screen allows you to select whether the zero
calibration is to be performed automatically or manually. Use the

∆∇

arrow

keys to toggle between

AUTO and MAN zero settling. Stop when AUTO

appears, blinking, on the display.

Select zero

mode: AUTO

Press

Enter

to begin zeroing.

#### % O2 — N2

Slope=##### C—Zero

The beginning zero level is shown in the upper left corner of the display. As

the zero reading settles, the screen displays and updates information on

Slope=

in percent/second (unless the Slope starts within the acceptable zero range and
does not need to settle further). The system first does a course zero, shown in
the lower right corner of the screen as

C—Zero, for 3 min, and then does a fine

zero, and displays

F—Zero, for 3 min.