4 Operation

Model 2000A-EU

4-20

Teledyne Analytical Instruments

The autoranging feature can be overridden so that analog output stays on a

fixed range regardless of the contaminant concentration detected. If the concen-
tration exceeds the upper limit of the range, the DC output will saturate at 1 V dc
(20 mA at the current output).

However, the digital readout and the RS-232 output of the concentration

are unaffected by the fixed range. They continue to read beyond the full-scale
setting until amplifier saturation is reached. Below amplifier saturation, the
overrange readings are accurate UNLESS the application uses linearization over
the selected range.

The concentration ranges can be redefined using the

Range

function

Manual screen, and the application gases can be redefined using the

System

function, if they are not already defined as necessary.

CAUTION: Redefining applications or ranges might require

relinearization and/or recalibration.

To setup automatic ranging:

Press

Range

key to start the Range function.

Select range mode

Range mode : AUTO

Note: If all three ranges are currently defined for different applica-

tion gases, then the above screen does not display (because
mode must be manual). Instead, the VFD goes directly to the
following screen.

If above screen displays, use the

∆∇

arrow keys to Select

AUTO, and

press

Enter.

Select auto ranging

Gas use: O2 — N2

Use the

∆∇

arrow keys to change the application (gas use:).

Press

Escape

to return to the previous screen to select or define another

range.

Press

Enter

to return the to the

Analyze

function.

4.6.3 Precautions

The Model 2000 allows a great deal of flexibility in choosing ranges for

automatic range switching. However, there are some pitfalls that are to be
avoided.