Thermal Conductivity Analyzer

Operation 4

4-1

Teledyne Analytical Instruments

Operation

4.1

Introduction

Although the Model 2000 is usually programmed to your application at the

factory, it can be further configured at the operator level, or even, cautiously,
reprogrammed. Depending on the specifics of the application, this might include
all or a subset of the following procedures:

Setting system parameters:

Establish a security password, if desired, requiring Operator
to log in.

Establish and start an automatic calibration cycle, if desired.

Routine Operation:

Calibrate the instrument.

Choose autoranging or select a fixed range of analysis.

Set alarm setpoints, and modes of alarm operation (latching,
failsafe, etc).

Program/Reprogram the analyzer:

Define new applications.

Linearize your ranges.

If you choose not to use password protection, the default password is

automatically displayed on the password screen when you start up, and you
simply press

Enter

for access to all functions of the analyzer.

4.2

Using the Data Entry and Function

Buttons

Data Entry Buttons: The < > buttons select options from the menu

currently being displayed on the VFD screen. The selected option blinks.

When the selected option includes a modifiable item, the

∆∆∆∆∆∇

arrow buttons

can be used to increment or decrement that modifiable item.