beautypg.com

Teledyne 2000A-EU - General purpose thermal conductivity analyzer User Manual

Page 25

background image

Teledyne Analytical Instruments

Thermal Conductivity Analyzer

Installation 3

3-3

All operator controls are mounted on the control panel, which is

hinged on the left edge and doubles as the door that provides access to the
sensor inside the instrument. The door is spring loaded and will swing open
when the button in the center of the latch (upper right corner) is pressed all
the way in with a narrow gauge tool (less than 4.5 mm wide), such as a
small hex wrench or screwdriver Allow clearance for the door to open in a
90-degree arc of radius 19.3 cm. See Figure 3-2.

Figure 3-2: Required Front Door Clearance

3.3

Electrical Connections (Rear Panel)

Figure 3-3 shows the Model 2000A-EU rear panel. There are connec-

tors for power, digital communications, and both digital and analog con-
centration output.

Figure 3-3: Rear Panel of the Model 2000A-EU