Luminex MasterPlex CT Tutorial Manual User Manual
Page 7
http://www.miraibio.com
Plex
CT
7
4. After you have placed the appropriate alignment tools into the well, press the
Location button and specify the well where you have placed them.
5. Now loosen the bolt that holds the probe (refer to the picture in the Adjust Probe
window) with the hex driver that is provided within the left panel of the Luminex
system where the heater block is stored.
6. Click on the Probe Down button to lower the probe arm. The probe height will
get adjusted by the spacers in the well.
7. Using the thumb wheel, lower the sample probe until it rests on, but does not
apply pressure to, the top of the alignment disk or sphere.
8. Tighten the bolt back up with the hex driver.
9. Click the Probe Up button to raise the sample probe.
10. Manually open the XYP door, and click on the Probe Down button again.
11. Watch the black XYP tray to see if it moves. If it moves, repeat this process until
no movement is observed.
12. Click the Probe Up button to raise the sample probe.
13. Replace the plastic shield that covers the sample probe area.