Probe Height Adjustment
Being able to adjust the probe height is crucial for any Luminex user as it is the cause
for many known issues such as:
Sample empty errors
Slow event rates
Air in the system
Probes being stuck in the down position
Actuator errors in the software
Adjusting the probe height must also be performed after any of the following events:
A new type of plate is being used
Probe has been removed (ex. For the declogging procedure)
Luminex system has been moved
You can find the Probe
Height launch button in
the left task bar under the
Expert Tools menu.