Luminex MasterPlex CT Tutorial Manual User Manual
Page 6
http://www.miraibio.com
Plex
CT
6
Thumbwheel
Bolt
The initial Adjust Probe window gives detailed instructions on adjusting the probe
height. To adjust the vertical height of the sample probe:
1. Remove the clear plastic shield that covers the sample probe area.
2. Press the Eject button and place a plate in the plate holder.
3. Place the appropriate alignment tool in the plate:
For a standard plate with flat-bottom wells – stack two of the larger
(5.08 mm diameter) alignment discs together and place them into a
selected well
For a filter bottom plate – stack three of the larger (5.08 mm diameter)
alignment discs together and place them into a selected well
For a half-volume plate with flat-bottom wells – stack two of the
smaller (3.35 mm diameter) alignment discs together and place them into
the selected well
For a round-bottom (U-bottom) plate – stack two of the smaller (3.35
mm diameter) alignment discs and place them into a selected well
For a plate with conical wells – place one alignment sphere into a
selected well
Master