1 volume calibration using the liquid sensor, 1 reset the calibration – Hach-Lange SIGMA SD900 Refrigerated Sampler and All Weather Refrigerated Sampler User Manual User Manual
Page 58

56
Sampler operation
4.8.1 Volume calibration using the liquid sensor
Use the volume calibration with the liquid sensor enabled to adjust
sample volumes slightly.
Procedure:
1. Make sure the liquid sensor is enabled by selecting
PROGRAM
SETUP
>
MODIFY SELECTED
>
LIQUID SENSOR
>
ENABLE
.
2. Press
VOLUME CALIBRATION
and select
CALIBRATION
.
3. Place the tubing inside the sampler cabinet that leads to the
sample bottle(s) into a graduated cylinder. Place the intake
tube into the sample source or tap water.
Note: Use the sample source for best accuracy.
4. Select
START
.
5. The pump will purge the intake tube and then collect the
sample volume that is specified in the sampling program. The
pump will then purge the intake tube.
To halt the calibration at any time, press the
STOP
key.
6. When the sample has been collected, select
DONE
. Compare
the volume collected in the graduated cylinder with the sample
volume that is entered in the program setup menu.
Select
REPEAT
to repeat the volume collection if necessary.
7. If the volume that was collected is different from the sample
volume in the current program, enter the volume that was
actually collected. Press
ENTER
. The sensor is then adjusted to
accurately measure all programmed volumes.
To verify sample volume, collect a grab sample (
8. Select
START
to start the sampling program or
CANCEL
to exit
the volume calibration menu.
4.8.1.1 Reset the calibration
Use the reset cal option to restore the volume calibration to the
default factory settings.
Procedure:
1. Make sure the liquid sensor is enabled by selecting
PROGRAM
SETUP
>
MODIFY SELECTED
>
LIQUID SENSOR
>
ENABLE
.
2. To undo the calibration, press
VOLUME CALIBRATION
and
select
RESET CAL
.
3. The calibration will be reset to the default calibration and the
message “value has been reset to 0” will be shown.