Appendix A Wiring for non-Sigma flow meters
CAUTION
Only qualified personnel should conduct the tasks described
in this section of the manual.
Prerequisites:
•
Multi-purpose half cable
•
Auxiliary adapter cable (6-pin to 7-pin adapter)
•
Optional splitter (
) to provide additional
connections. Two or more splitters may be connected in series.
Procedure:
Use the auxiliary connector diagram in
and wiring
to wire the flow meter to the controller.
Figure 41 Auxiliary connector (7-pin)
Table 9 Half-cable wiring information for non-Sigma flow meters
Pin
Signal
description
Wire
color
1
Purpose
Rating
A
12 VDC
power
output
White
Power supply positive output. Powers
an external device or flow meter. Must
be used in conjunction with pin B
(common).
This output is protected by a thermal current limiter.
12 VDC nominal output at 1 A maximum when using
battery 8754400 to power the sampler.
15 VDC regulated output at 0.5 A maximum when
using AC power to power the sampler.
B
Common
Blue
Power supply negative return. When
using AC power, this pin is connected to
earth ground
2
.