Basic user manual, Specifications, General information – Hach-Lange POLYMETRON 8350_8351 Basic User Manual User Manual
Page 3: Safety information, English

Specifications
The 8350 pH and 8351 Redox probes are dedicated for process measurements. The pH probes
incorporate impedance control and a temperature sensor in contact with the sample.
8350.4
8350.5
8351
1 - Glass
2 - Tube containing PT100
3 - Platinum ring
4 - PTFE Junction
A special model (8350.3) is compatible with all the standard accessories and is designed for use in
hydrofluoric acid.
8350.4
8350.5
8351
Application
Measuring pH at high
temperatures
Measuring pH in a
waste water
environment
Measuring redox
Junction type
Flat, PTFE
Flat, PTFE
Flat, PTFE
Body material
PPS
CPVC
PPS
Measurement range
0-14 pH
0-12 pH
± 1500 mV
Maximum temperature (°C)
110
80
110
Maximum pressure
10 bar
10 bar
10 bar
Drift (/ week)
< 3 mV
< 3 mV
< 3 mV
Offset @ pH7
± 20 mV
± 20 mV
N/A
Slope
56-61 mV/pH
56-61 mV/pH
N/A
Reference impedance @ 25°C
< 50 Kohm
< 50 Kohm
< 50 Kohm
Glass impedance @ 25°C
150-500 Mohm
150-500 Mohm
Platinum
General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.
Safety information
Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention
to all danger and caution statements. Failure to do so could result in serious injury to the operator or
damage to the equipment.
English 3