Table 3 – Fluid Components International OEM Flow Meter User Manual

Page 3

background image

OEM MASS FLOW

FLUID COMPONENTS INTERNATIONAL LLC

The top level of the menu is shown below. Enter the large letter in the tables below to activate a command.

Doc. No. 06EN003329 Rev. B

“Z” Flow Units Set-up and Scaling

Units

Select

E=English

M=Metric

Select

F= SFPS

M=SMPS

or

C = SCFM

N = NCMH

or

H = SCFH

O = NCMM

or

L = LBS/H

K = KG/H

For Volumetric or Mass Flow

Select

R = Round pipe or duct

or

S = Square duct

Set

Diameter or Wide X High (in inches or mm)

Set

Max = Maximum flow rate (span)

Set

Zero = Minimum flow rate (zero)

Table 3

“V” Output Configuration Set-up

Analog out

Select

1

2

3

4

4-20mA out

Flow

Flow

Temp

Temp

0-10 Vout

Temp

Flow

Flow

Temp

Pulse out

Select

1

2

3

4

Source out

Pulse

Pulse

Alarm0

Alarm0

Set

Factor

Factor

Set pt.0

Set pt.0

Set

Period

Period

State0

State0

Set

State0

State0

Sink

Pulse

Alarm1

Pulse

Alarm1

Set

Set pt.1

Factor

Set pt.1

Set

State1

State1

Period

State1

State1

Table 4

Table 2

User may exit command at any time entering
“Q” [ENTER] in menus: D, K, L, V, W, or Z.

C

Calibration Information
Display only: A/D, Delta-R, Ref-R data values

D

Diagnostics
Display only: Check-out of functional conditions

K

Factory Calibration Settings
Display only: Cal. parameters, i.e. linearization
and temperature compensation coefficients.

L

Calibrate Outputs

Factory Access Only

Y

Command Line Interface Factory Access Only

R

Factory Reset
Replaces user data with factory calibration data

Diagnostics and Factory Settings

Analog out
Output Mode
Selected

4-20mA: Flow
0-10V: Temp

Change? (Y/N)>

4-20mA: Flow
0-10V: Temp

Enter 1 to make
the selection__

4-20mA: Flow
0-10V: Flow

Enter 2 to make
the selection__

4-20mA: Temp
0-10V: Flow

Enter 3 to make
the selection__

4-20mA: Temp
0-10V: Temp

Enter 4 to make
the selection__

Display comes up to last setting saved
and stays for 2 seconds.
If N or [ENTER], proceed to Pulse out.
If Y, display moves to selection
options and asks for confirmation. If
you miss the option, select [Enter]
repeatedly to loop around.

Pulse out
Pulse Out
Selected

Source: Pulse
Sink: Pulse

Change? (Y/N)>

Source: Pulse
Sink: Pulse

Enter 1 to make
the selection #__

Source: Pulse
Sink: Alarm1

Enter 2 to make
the selection #__

Source: Alarm0
Sink: Pulse

Enter 3 to make
the selection #__

Source: Alarm0
Sink: Alarm1

Enter 4 to make
the selection #__

PFactor: 1.000
Change? (Y/N)>

if yes
Enter new factor: ____

Sample Period: 1 second

Change? (Y/N)>
if yes
Enter new Sample Period: ____

If alarm is a selected output
Set point1: 000

Set points are in the
same units as the flow
or temp.

Change? (Y/N)>

if yes
Enter new set point: ____

Resume normal operation

Source state: 0
Change to 1?>

Sink state: 1
Change to 0?>

T

Normal Operation Mode

All outputs active
F

K Factor (1 is default)

Z

Flow Units Set-up
Select flow units (4 English, 4 Metric),
Pipe diameter and span/zero settings

W Totalizer Mode

Disable/Enable internal totalizer

V

Output Configuration Selection
Select one of four output configurations: pulse
and/or alarm, plus pulse factor and/or set point

N

Warm Reset
Re-initialize instrument without removing power

Table 1

Customer Settings

This page is subject to proprietary rights statement on last page

Note: Changing units requires rescaling the unit (set new zero and span).