beautypg.com

Fluid Components International ST98 Manual Installation User Manual

Page 6

background image

FLUID COMPONENTS, INTL

CHAPTER 2 - INSTALLATION

Model ST98 Flow Meter

2 - 6

Doc. No. 06EN003291 Rev. A

Figure 2-3. Circuit Board Placement

9.

For remote instruments only: The flow element wires should be routed through the 3/4 inch NPT port for the
flow element as shown in Figure 2-3. Connect the flow element wires to TS2 on the electronics assembly
according to Figure 2-5. Connect the cable shield to HTR RTN. Leave the other end of the shield floating.
A 14 AWG ground wire should also be routed between the enclosures (wire terminals are supplied in the kit).

Note:

Connecting the shield in any other way will decrease the accuracy of the instrument. See Figure 2-4

for the wiring diagram.

Caution:

Be sure an earth ground wire is connected to the ground terminal (see Figure 2-4). On a remote
configuration, connect an earth ground wire to the ground screw in the local enclosure. This is for
the purpose of safety.

10. If a wire comes loose from the instrument during installation, refer to Chapter 5 - Troubleshooting for a

complete instrument wiring diagram.

Figure 2-4. Customer Connection Board

POWER

AND

SIGNAL

CONNECTION

BOARD

3/4 INCH NPT PORT (2 PLACES)

OPTIONAL

DISPLAY

GLASS

3/4 INCH NPT

PORT FOR

FLOW ELEMENT

ELECTRONICS

ELECTRONICS COVER
(NO USER ACCESS
NECESSARY FOR
INTEGRAL INSTRUMENTS)

CUSTOMER

CONNECTION

COVER

OPTIONAL DISPLAY

P1

TS1

TS4

TS3

TIE DOWN
BRACKET

CABLE TIES

FERRITE BEAD