beautypg.com

Example – Fluid Components International ST98 Manual Operation User Manual

Page 15

background image

CHAPTER 3 - OPERATION

FLUID COMPONENTS INTL

Doc. No. 06EN003291 Rev. A

3 - 15

Model ST98 Flow Meter

any turn down ratios lower than 10:1. If an attempt is made to enter a lower value, an error will be displayed and
an opportunity to reenter is given.

Example:

Note:

Pressing [Q] [ENTER] at any time will exit the menu. All data will return to the original values
prior to entering this menu.

Velocity (English)

Pressing [Z] [ENTER] will display “E for English or” followed by “M for Metric”. Press an [E] [ENTER]
and the next prompt is “F SFPS, C SCFM” followed by “or L LBS/HR”. Press an [F] [ENTER] and the next
prompt is “Max = xxx.xxxxxx” followed by “Change F.S. ?<”. Press [Y] [ENTER] and enter a new value, or
else press [N] [ENTER] or [ENTER]. The next prompt is “Zero = x.xxxxxx” followed by “Change it ?<”.
Press [Y] [ENTER] and enter a new value or else press [N] [ENTER] or [ENTER]. The next prompt is
Save ?<”. Press [Y] [ENTER] and the new values are saved to the EEPROM. The last prompt is
Input mode ?<”.

Volume (English)

Pressing [Z] [ENTER] will display “E for English or” followed by “M for Metric”. Press an [E] [ENTER]
and the next prompt is “F SFPS, C SCFM” followed by “or L LBS/HR”. Press [C] [ENTER] and the next
prompt is “R round duct or” followed by “S rectangular <”. Press [R] [ENTER] and the next prompt is
Diameter: inches” “x.xxxxxx” followed by “Change it ?<”. Press [Y] [ENTER] and enter a new value or
else press [N] [ENTER] or [ENTER]. If [S] [ENTER] is pressed, the next prompt is
Inches Wide” “x.xxxxxx” followed by “Change it ?<”. Press [Y] [ENTER] and enter a new value or else
press [N] [ENTER] or [ENTER]. The next prompt is “Inches High”. “x.xxxxxx” followed by
Change it ?<”. Press [Y] [ENTER] and enter a new value or else press [N] [ENTER] or [ENTER]. The
next prompt is “Max = xxx.xxxxxx” followed by “Change F.S. ?<”. Press [Y] [ENTER] and enter a new
value or else press [N] [ENTER] or [ENTER]. The next prompt is “Zero = x.xxxxxx” followed by
Change it ?<”. Press [Y] [ENTER] and enter a new value or else press [N] [ENTER] or [ENTER]. The
next prompt is “Save ?<”. Press [Y] {ENTER] and the new values are saved. The last prompt is
Input mode ?<”.

Mass Flow (English)

Pressing Z [ENTER] will display “E for English or” followed by “M for Metric”. Press [E] [ENTER] and
the next prompt is “F SFPS, C SCFM” followed by “or L LBS/HR”. Press [L] [ENTER] and the next
prompt is “R round duct or” followed by “S rectangular <”. Press [R] [ENTER] and the next prompt is
Diameter: inches” “x.xxxxxx” followed by “Change it ?<”. Press [Y] [ENTER] and enter a new value or
else press [N] [ENTER] or [ENTER]. If [S] [ENTER] is pressed, the next prompt is
Inches Wide” “x.xxxxxx” followed by “Change it ?<”. Press [Y] [ENTER] and enter a new value or else
press [N] [ENTER] or [ENTER]. The next prompt is “Inches High”, “x.xxxxxx” followed by
Change it ?<”. Press [Y] [ENTER] and enter a new value or else press [N] [ENTER] or [ENTER]. The next
prompt is “Max = xxx.xxxxxx” followed by “Change F.S. ?<”. Press [Y] [ENTER] and enter a new value or
else press [N] [ENTER] or [ENTER]. The next prompt is “Zero = x.xxxxxx” followed by “Change it ?<”.
Press [Y] [ENTER] and enter a new value or else press [N] [ENTER] or [ENTER]. The next prompt is
“Save ?<”. Press [Y] [ENTER] and the new values are saved. The last prompt is “Input mode ?<”.

Velocity (Metric)

Pressing [Z] [ENTER] will display “E for English or” followed by “M for Metric”. Press an [M] [ENTER]
and the next prompt is “M SMPS, N NCMH” followed by “or K KG/HR”. Press an [M] [ENTER] and the
next prompt is “F.S. = xxx.xxxxxx” followed by “Change it ?<”. Press [Y] [ENTER] and enter a new value or
else press [N] [ENTER] or [ENTER]. The next prompt is “Zero = x.xxxxxx” followed by “Change it ?<”.
Press [Y] [ENTER] and enter a new value or else press [N] [ENTER] or [ENTER]. The next prompt is
Save ?<”. Press ‘Y’ and the new values are saved. The last prompt is “Input mode ?<”.

Volume (Metric)