Alarms – Burkert Type 8056 User Manual
Page 30
SE56
30
CODE
ANOMALIE DESCRIPTION
ACTION TO TAKE
0001
problem with watch-dog circuit
ADDRESS TO SERVICE
0002
Wrong work data configuration in eeprom
0004
Wrong safety data configuration in eeprom
0008 Defective
eeprom
0010
Defective keyboard (one or more keys are pushed
during the test)
0020
Power supply voltage (+3.3) is out of range
0040
Power supply voltage (+13) is too low (<10V)
0080
Power supply voltage (+13) is too high (>14V)
0200
Timeout calibration input (broken input circuit)
0400
Input stage gaining is out of range
Check the status cables connecting the transmitter to the sensor, the
grounding connections or the possible presence of strong and
anomalous noise sources
0800
Interruption on the coils circuit
Check the status cables connecting the transmitter to the sensor
0C00
Cumulative alarm 0800 + 0400
See single code
ALARMS
Messages, causes and corrective actions
Anomalies, codes and remedies
MESSAGE
CAUSE
ACTION TO TAKE
NO ALARMS
All works regularly
//
MAX ALARM
The flow rate is higher than the maximum threshold set
Check the maximum flow rate threshold set and
the process conditions
MIN ALARM
The flow rate is lower than the minimum threshold set
Check the minimum flow rate threshold set and
the process conditions
FLOW RATE >FS
The flow rate is higher than the full scale value set on the
instrument
Check the full scale value set on the instrument
and the process conditions
PULSE/FREQ>FS
The pulse generation output of the device is saturated and can
not generate the sufficient number of pulses
Set a bigger volume unit or, if the connected
counting device allows it, reduce the pulse
duration value
EMPTY PIPE
The measuring pipe is empty or the detection system has not
been properly calibrated
Check whether the pipe is empty, or repeat the
empty pipe calibration procedure
INPUT NOISY
The measure is strongly effected by external noise or the
connecting cable from transmitter to the sensor may be broken
Check the status cables connecting transmitter to
sensor, the devices grounding connections or the
possible presence of noise sources
EXCITATION
FAIL
The coils or the cable connecting the sensor are interrupted
Check the connecting cables to the sensor
CURR. LOOP
OPEN
The 4-20mA output on board or the optional one are not
correctly closed on a valid load
Verify the load is applied to the output (max 1000
).
To disable the alarm, set the “mA VAL.FAULT”
value (menu alarm) to 0.
P.SUPPLY FAIL
Power supply different from that indicated on the label
Verify that the power supply is as indicated on the
label