Input on/off operation, Digital input, Auto-calibration reset totalizers block totalizers – Burkert Type 8056 User Manual
Page 10: Range change, External power supply internal power supply
SE56
10
0-1.5V
3-40V
AUTOCALIB. OFF
T
RESET
0-1.5V
3-40V
Tmin = 100ms
BLOCK
0-1.5V
3-40V
T
0-1.5V
3-40V
SCALE 1
SCALE 2
<18V
18-30V
3/40 Vdc (ON)
0/1.5 Vdc (OFF)
5 (+)
6 (-)
10 K
5
6
20
0
15
+24
10 K
DIGITAL INPUT
INPUT ON/OFF OPERATION
Auto-calibration
Reset totalizers
Block totalizers
Block totalizers
Totalizers active
Necessary conditions for enabling the function
Necessary conditions for enabling the function
Necessary conditions for enabling the function
Tmin
POS. 5.7 ENABLED (Autozero calibration external
command)
POS. 5.1 to 5.4 ENABLED (
partial positive or reverse
flow totalizers reset enable).
POS. 5.5 ENABLED (
totalizers counting lock command)
Necessary conditions for enabling the function
Range change
10Hz
20Hz
50Hz
80Hz
150Hz
220ms
110ms
45ms
30ms
15ms
T must be > Tmin
SAMPLE RATE
Tmin
SAMPLE RATE
Tmin
External power supply
Internal power supply
POS. 5.8 ENABLED (range change)
Measure lock
Measure blocked
Measure
Necessary condition for enabling the function
POS. 5.6 ENABLED (measure lock)