beautypg.com

Brooks, Model 5851e – Brooks Instrument 5851E User Manual

Page 29

background image

3-9

Brooks

®

Model 5851E

Section 3 Operation

Installation and Operation Manual
X-TMF-5851E-MFC-eng
Part Number: 541B104AAG
November, 2008

Example;
What is the percent of full scale error when full scale is equal to 100
slpm?
Measured flow rate = 48.5 slpm
Desired flow rate = 50.0 slpm

(48.5 - 50)

Full scale error = 100

= -1.5%

100

i. Calculate the TP2 correction voltage:

(error recorded in Step h) x 0.450 Volts

Example:
Error = -1.5%
TP2 correction voltage = -1.5 x 0.450 = -0.675 Volts
New TP2 voltage = 0 Volts + (-0.675) = -0.675 Volts

j. Set the command potentiometer for 100% flow (5.000 V). Connect the

DVM positive lead to TP2 and the negative lead to TP4.

k. Adjust the linearity potentiometer for an output equal to the new calcu-

lated TP2 voltage.

l. Repeat Steps f, g and h.

1. If the error recorded in Step h is less than 0.5%, then the calibration

procedure is complete.

2. If the error is greater than 0.5% set the command potentiometer for

100% (5.000 V). Connect the DVM positive lead to TP2 (linearity
voltage) and the negative lead to TP4 (circuit common). Calculate a
new TP2 voltage as follows:

error

measured

New TP2 voltage =

recorded in

x

0.450 V

+

TP2

Step i

voltage

Example:

Controller error = 0.7%
Measured TP2 voltage = -0.567 Volts
TP2 correction = 0.7 x 0.450 = 0.315 Volts
New TP2 correction = 0.315 + (-0.567) = 0.252 Volts

Adjust the linearity potentiometer for an output equal to the new TP2
voltage and then repeat Steps f, g and h.

Note: The voltage at TP2 can range from -10 to +3 Volts, however, it
is recommended that this voltage stay between -2.5 and +2.5 Volts
for proper operation. If the recommended voltage range is exceeded
the desired accuracy and/or signal stability may not be achieved. If
one of the limits is reached, check the restrictor sizing. Refer to
section 4-7.

3-5 Response

Fast Response Adjustment
Two methods of adjusting the step response of the Model 5851E mass
flow controllers can be used. No. 1 describes a method that will get the
step response close to optimum quickly and without any flow measuring
equipment. This method should be used when the response time of the