beautypg.com

Solidsense ii, Series – Brooks Instrument SolidSense II User Manual

Page 25

background image

Installation and Operation Manual
X-PR-SolidSense II-PT-eng
Part Number 541B160AAG
July, 2011

SolidSense II

®

Series

2-9

Section 2 Installation

Position the SolidSense II pressure transducer with the electrical connector
facing pointing upward. Using a small, flat-bladed screw driver, turn zero
adjustment screw in a clockwise direction to increase the signal. Turn the
screw in a counterclockwise direction to decrease the signal.

Figure 2-9 Zero Adjustment Screw Location on the (Standard
SolidSense II pressure tranducers)

Zero Adjustment Screw

2-10 Zero Adjustment

The SolidSense II pressure transducer is factory calibrated and does not
normally need a field adjustment. If a field adjustment becomes necessary,
adjust the zero as noted below.

2-10-1 Zero Adjustment (SolidSense II Standard Pressure Transducers)

On standard SolidSense II pressure transducer models, the zero
adjustment screw is located on the body as shown in Figure 2-8.