0x6b read the bar code information – B&K Precision 8500-8526 - Manual User Manual
Page 69

The operation state register's bit meanings are:
Bit
Meaning
0
Calculate the new demarcation coefficient
1
Waiting for a trigger signal
2
Remote control state (1 means enabled)
3
Output state (1 means ON)
4
Local key state (0 means not enabled, 1 means enabled)
5
Remote sensing mode (1 means enabled)
6
LOAD ON timer is enabled
7
Reserved
The demand state register's bit meanings are:
Bit
Meaning
0
Reversed voltage is at instrument's terminals (1 means yes)
1
Over voltage (1 means yes)
2
Over current (1 means yes)
3
Over power (1 means yes)
4
Over temperature (1 means yes)
5
Not connect remote terminal
6
Constant current
7
Constant voltage
8
Constant power
9
Constant resistance
0x6A Get product's model, serial number, and firmware version
Byte
offset
Meaning
3 to 7
ASCII model information
8
Low byte of firmware version number
9
High byte of firmware version number
10 to 19 Instrument's serial number in ASCII
20-24
Reserved
0x6B Read the bar code information
Note: the bar code information is to be interpreted as ASCII data.
8500 DC Load Series
Version: 030614