Pulse transient operation – B&K Precision 8500-8526 - Manual User Manual
Page 36

Keys
Display
I-set
Standard display of voltage and current (or, you'll be asked to enter a
current value).
Shift + S-Tran
LEVEL A= X.XXXA
You are being prompted for the first current value.
Press the 5 key for 5 A.
Enter
Enters the 5 A value, then displays
WIDTH A = X.XMS
. You're being
prompted for the duration of the 5 A load -- press the 3 key for 3 ms.
Enter
Enters the 3 ms value, then displays
LEVEL B= X.XXXA
. You are being
prompted for the second current value. Press the 1 and 0 keys to enter
10 A.
Enter
Enters the 10 A value, then displays
WIDTH B = X.XMS
. You're being
prompted for the duration of the 10 A load -- press the 2 key for 2 ms.
Enter
Enters the 2 ms duration, then displays
:CONTINUOUS
,
:PULSE
, or
:TOGGLED
. Use the
△ or ▽ arrow keys to display
:CONTINUOUS
.
Enter
Returns to the standard display of voltage and current.
To activate this continuous transient condition, press Shift + Tran, then press On/Off. The load will
begin switching between the two Tran values with the timing you entered.
Pulse transient operation
In pulse transient operation, the load operates at the A value that has been entered until a trigger is
received. At the trigger, the load switches to the B value and stays at that level for the B timing
value. Then the load switches back to the A value and stays there until another trigger is received.
Here's an example:
10A
5A
Pulsed Transient Operation
10ms
TWD
10ms
TWD
TRIG
TRIG
Here are the keystrokes needed to set up this example:
Keys
Display
I-set
Standard display of voltage and current (or, you'll be asked to enter a
current value).
Shift + S-Tran
LEVEL A= X.XXXA
You are being prompted for the first current value.
Press the 5 key for 5 A.
Enter
Enters the 5 A value, then displays
WIDTH A = X.XMS
. You're being
prompted for the duration of the 5 A load. In the pulse mode of operation,
8500 DC Load Series
Version: 030614
Page 36 of 77