0x45 read one of the step's power and time values, 0x48 set the list file name, 0x49 read the list file name – B&K Precision 8500-8526 - Manual User Manual
Page 64: 0x4a set the memory partitioning for storing lists

0x45 Read one of the step's power and time values
Byte
offset
Meaning
3 to 4
2 byte little-endian integer specifying which step number in the list
5 to 8
4 byte little-endian integer specifying the power in units of 1 mW
9 to 10
2 byte little-endian integer specifying the step timing in units of 0.1 ms
11-24
Reserved
0x46 Set one of the step's resistance and time values
Byte
offset
Meaning
3 to 4
2 byte little-endian integer specifying which step number in the list
5 to 8
4 byte little-endian integer specifying the resistance in units of 1 m
9 to 10
2 byte little-endian integer specifying the step timing in units of 0.1 ms
11-24
Reserved
0x47 Read one of the step's resistance and time values
Byte
offset
Meaning
3 to 4
2 byte little-endian integer specifying which step number in the list
5 to 8
4 byte little-endian integer specifying the resistance in units of 1 m
9 to 10
2 byte little-endian integer specifying the step timing in units of 0.1 ms
11-24
Reserved
0x48 Set the list file name
Byte
offset
Meaning
3 to 12
List file name (ASCII characters)
13-24
Reserved
0x49 Read the list file name
Byte
offset
Meaning
3 to 12
List file name (ASCII characters)
13-24
Reserved
0x4A Set the memory partitioning for storing lists
8500 DC Load Series
Version: 030614
Page 64 of 77