Byte
offset
Meaning
5
Upper low byte of resistance.
6
Upper high byte of resistance.
7-24
Reserved
0x32 Set CC mode transient current and timing
Byte
offset
Meaning
3 to 6
Value A of current in units of 0.1 mA. Little-endian 4 byte number.
7 to 8
Time for A current in units of 0.1 ms. Little-endian 2 byte number.
9 to 12
Value B of current in units of 0.1 mA. Little-endian 4 byte number.
13 to 14 Time for B current in units of 0.1 ms. Little-endian 2 byte number.
15
Transient operation:
0 is CONTINUOUS
1 is PULSE
2 is TOGGLED
16-24
Reserved
0x33 Read CC mode transient parameters
Byte
offset
Meaning
3 to 6
Value A of current in units of 0.1 mA. Little-endian 4 byte number.
7 to 8
Time for A current in units of 0.1 ms. Little-endian 2 byte number.
9 to 12
Value B of current in units of 0.1 mA. Little-endian 4 byte number.
13 to 14 Time for B current in units of 0.1 ms. Little-endian 2 byte number.
15
Transient operation:
0 is CONTINUOUS
1 is PULSE
2 is TOGGLED
16-24
Reserved
0x34 Set CV mode transient voltage and timing
Byte
offset
Meaning
3 to 6
Value A of voltage in units of 1 mV. Little-endian 4 byte number.
7 to 8
Time for A voltage in units of 0.1 ms. Little-endian 2 byte number.
9 to 12
Value B of voltage in units of 1 mV. Little-endian 4 byte number.
13 to 14 Time for B voltage in units of 0.1 ms. Little-endian 2 byte number.
15
Transient operation:
8500 DC Load Series