Static Recording

This mode is used for recording maximum, minimum, and
average capacitance values. It captures only stable values
and updates to memory. Values that result in displaying
OL” or that are below 10 counts are not recorded. This
mode is useful for determining the range of values a
component falls into.

Enable Static Recording

Press the

button to enter the static recording mode.

The display should indicate “MAX AVG MIN
simultaneously. This indicates the meter is in static
recording mode and recording is performed immediately.

Note: In static recording mode, the auto power off function
will be disabled and the “@OFF” indicator will be turned off
automatically. It will be re-enabled with its previous settings
once meter exits static recording mode.

Using Static Recording


There are four different modes that can be selected in
static recording. These modes can be changed with

each

button press. With each press of the

button, the modes will cycle through as

followed: :