3.2 INSTRUMENT

SPECIFICATIONS

Electrical Specifications

(See tables 2)

NOTE: (1) The test leads should be as short as possible to

minimize the measurement error.

(2) For best accuracy, zero adjustment should be

performed appropriately before testing.

Table 2.
CAPACITANCE

Range

*Accuracy Resolution

Test Condition

200pF

0.1pF

2nF

1pF

20nF

10pF

200nF

100pF

2µF

1000pF

Parallel Mode

1KHz, 0.5Vrms

20µF

0.01µF

Series Mode 120Hz, 1mArms

200µF

1%+2

0.1µfF

2mF

1µF

20mF

2%+10

10µfF

Series Mode

120Hz, 10mArms

*Accuracy is ± ( % of reading+number of digits ) When D≦0.1

DISIPATION FACTOR

Range

Accuracy

200pF 2nF≦Cx≦2µF

2µF<Cx≦2mF

20mF

0 ~ 1.999

N.S.

Cx

2000

20

%

2

+

+

N.S.

Cx

2000

10

%

1

+

+

Accuracy is ± ( % of reading+number of digits ).

Cx is capacitance readout in counts.

Accuracy is applied when C is from 20 to 100% of full scale range in series

mode measurements.

Parallel Mode

fCpRp

Dc

π

2

1

=

Cp

Rp