7 make capacitance measurements – B&K Precision 393 - Manual User Manual
Page 18
![background image](/manuals/320661/18/background.png)
18
Follow these steps for continuity testing:
1.
Connect the black test lead to the COM input.
2.
Connect the red test lead to the VΩHz%ns
input.
3.
Connect the two leads together to verify that the
continuity function is working properly. The
instrument should have a continuous beep sound.
4.
Probe with the test leads to the DUT and take the
measured reading on display.
5.
If continuity is good, it will have a continuous beep
sound.
Figure 4 - Connection for Continuity Test
WARNING:
Do not apply more than 1000 V DC across the
terminals or they will be damaged.
2.7 Make Capacitance Measurements
Turn the rotary switch to the corresponding symbol for the
measurement you would like to perform.
WARNING:
Fully discharge the capacitor before connecting it
to any of the inputs or it may damage the
instrument.
20 A
uAmA
COM
V.Ω
Continuity