Characteristics – Atec Tektronix-AWG7000 Series User Manual
Page 5
Arbitrary Waveform Generator — AWG7102 • AWG7101 • AWG7052 • AWG7051
Characteristics
Arbitrary Waveforms
Characteristic
AWG7102
AWG7101
AWG7052
AWG7051
Waveform Length
2 to 32,400,000 points (or 2
to 64,800,000 points, Option
01) in multiples of 64
Interleave:
2 to 64,800,000 points (or 2
to 129,600,000 points, Option
01) in multiples of 128
2 to 32,400,000 points (or 2 to 64,800,000 points, Option 01) in multiples of 64
Number of Waveforms
1 to 16,000
Sequence Length
1 to 4,000 steps
Sequence Repeat Counter
1 to 65,536 or infinite
Sequence Control
Repeat count, Wait for Trigger, Go-to-N, and Jump
Jump Mode
Synchronous and Asynchronous
Run Modes
Continuous
Waveform is iteratively output. If a sequence is defined, the sequence order and repeat functions are applied
Triggered
Waveform is output only once when an external, internal, GPIB, LAN, or manual trigger is received
Gated
Waveform begins output when gate is true and resets to beginning when false
Sequence
Waveform is output as defined by the sequence
Interleave operation
Up to 20 GS/s sample rate
(Option 06)
N/A
Clock Generator
Sampling frequency
10 MS/s to 10 GS/s
(10 GS/s to 20 GS/s at interleave)
10 MS/s to 10 GS/s
10 MS/s to 5 GS/s
Resolution
8 digits
Internal Clock
Accuracy
Within ±(1 ppm + Aging)
Aging: within ±1 ppm/year
Clock phase noise
< -90 dBc/Hz at 100 kHz offset
Internal Trigger Generator
Internal Trigger Rate
Range
1.0 μs to 10.0 s
Resolution
3 digits, 0.1 μs minimum
Skew Control Between Outputs
Range
-100 ps to +100 ps
N/A
-100 ps to +100 ps
N/A
Resolution
1 ps
N/A
1 ps
N/A
Skew accuracy
±(10% of setting +10 ps)
N/A
±(10% of setting +10 ps)
N/A
www.tektronix.com