beautypg.com

Wt3000 spec, Precision power analyzer wt3000, Wt3000 specifications – Atec Yokogawa-WT3000 User Manual

Page 12

background image

12

Precision Power Analyzer WT3000

(U1+U2)/2

(I1+I2)/2

P1+P2

S1+S2

U

Σ

[V]

I

Σ

[A]

P

Σ

[W]

S

Σ

[VA]

Q

Σ

[var]

Pc

Σ

[W]

WP

Σ

[Wh]

WP+

Σ

[Wh]

WP–

Σ

[Wh]

q

Σ

[Ah]

q+

Σ

[Ah]

q–

Σ

[Ah]

WQ

Σ

[varh]

WS

Σ

[VAh]

λΣ

Ø

Σ

[˚]

3 phase, 3 wire

3 phase, 3 wire

(3 voltage 3 current)

3 phase, 4 wire

P

Σ

2

+Q

Σ

2

TYPE1,

TYPE2

TYPE3

TYPE1

TYPE2

TYPE3

S1+S2+S3

Q1+Q2+Q3

Q1

+

Q2

+

Q3

Pc1

+

Pc2

+

Pc3

WP1

+

WP2

+

WP3

WP+1

+

WP+2

+

WP+3

WP-1

+

WP-2

+

WP-3

q1+q2+q3

q+1

+

q+2

+

q+3

q-1

+

q-2

+

q-3

P1+P2+P3

N

1

Σ

N

n=1

| Q

Σ

(n) |

×

Time

N

1

S

Σ

P

Σ

Σ

N

n=1

S

Σ

(

n

)

×

Time

cos

-

1

S

Σ

P

Σ

( )

3

2

3

3

(S1+S2)

(S1+S2+S3)

S

Σ

2

–P

Σ

2

Single-phase,
3 wire

(U1+U2+U3)/3

(I1+I2+I3)/3

Q1+Q2

Q1+Q2

Pc1

+

Pc2

WP1

+

WP2

WP+1

+

WP+2

WP–1

+

WP–2

q1

+

q2

q+1

+

q+2

q–1

+

q-2

Q

Σ

(

n

) is the nth reactive power

Σ

function , and N is the number of data updates.

S

Σ

(

n

) is the nth apparent power

Σ

function, and N is the number of data updates.

η

[%]

User-defined functions
F1–F20

Create equations combining measurement function symbols, and calculate up to
twenty numerical data.

Set a efficiency calculation up to 4

Waveform display items

Voltage and current from elements 1 through 4
Motor version torque and waveform of revolution speed

WT3000 SPEC

WT3000 Specifications

Inputs

Item

Specification

Input terminal type

Voltage
Plug-in terminal (safety terminal)
Current
• Direct input: Large binding post
• External sensor input: Insulated BNC connector

Input type

Voltage
Floating input, resistive potential method
Current
Floating input, shunt input method

Measurement range

Voltage

(rated value)

15 V, 30 V, 60 V, 100 V, 150 V, 300 V, 600 V, 1000 V (for crest factor
3)
7.5 V, 15 V, 30 V, 50 V, 75 V, 150 V, 300 V, 500 V (for crest factor 6)
Current (2A input element)
• Direct input:

5mA, 10mA, 20mA, 50mA, 100mA, 200mA, 500mA, 1A, 2A (for
crest factor 3)
2.5mA, 5mA, 10mA, 25mA, 50mA, 100mA, 250mA, 500mA, 1A
(for crest factor 6)

• External sensor input:

50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2 V, 5 V, and 10 V (for crest
factor 3)
25 mV, 50 mV, 100 mV, 250 mV, 500 mV, 1 V, 2.5 V, and 5 V (for
crest factor 6)
Current (30A input element)

• Direct input:

500 mA, 1 A, 2 A, 5 A, 10 A, 20 A, and 30 A (for crest factor 3)
250 mA, 500 mA,1 A, 2.5 A, 5 A, 10 A, and 15 A (for crest factor
6)

• External sensor input:

50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2 V, 5 V, and 10 V (for crest
factor 3)
25 mV, 50 mV, 100 mV, 250 mV, 500 mV, 1 V, 2.5 V, and 5 V (for
crest factor 6)

Input impedance

Voltage
Input resistance: Approx. 10 M

, input capacitance: Approx. 5 pF

Current (2A input element)
• Direct input: Approx. 500 m

+ approx. 0.07

µ

H

• External sensor input: Input resistance: Approx. 1 M

, input

capacitance: Approx. 40 pF

Current (30A input element)
• Direct input: Approx. 5.5 m

+ approx. 0.03

µ

H

• External sensor input: Input resistance: Approx. 1 M

, input

capacitance: Approx. 40 pF

Instantaneous maximum Voltage
allowable input

Peak value of 2500 V or RMS value of 1500 V, whichever is less.

(1s or less)

Current (2A input element)
• Direct input: Peak value of 9 A or RMS value of 3 A, whichever is

less.

• External sensor input: Peak value less than or equal to 10 times

the measurement range.

Current (30A input element)
• Direct input: Peak value of 150 A or RMS value of 50 A,

whichever is less.

• External sensor input: Peak value less than or equal to 10 times

the measurement range.

Continuous maximum

Voltage

allowable input

Peak value of 1600 V or RMS value of 1100 V, whichever is less.
Current (2A input element)
• Direct input: Peak value of 6 A or RMS value of 2.2 A, whichever

is less.

• External sensor input: Peak value less than or equal to 5 times

the measurement range.

Current (30A input element)
• Direct input: Peak value of 90 A or RMS value of 33 A, whichever

is less.

• External sensor input: Peak value less than or equal to 5 times

the measurement range.

Continuous maximum common mode voltage (50/60 Hz)

Voltage input terminals, current input terminals
1000 Vrms
External current sensor input connector
600 Vrms

Rated voltage to ground Voltage input terminals, current input terminals

1000 V
External current sensor input connector
600 V

Influence from common mode voltage

Apply 1000 Vrms with the voltage input terminals shorted and the
current input terminals open.
• 50/60 Hz:

±

0.01% of range or less

• Reference value up to 200 kHz
Voltage:

±

3/range

×

f% of range or less. However, 3% or less.

Current direct input and current sensor input:

±

(max. range/range)

Ч

0.001

Ч

f% of range or less.

However, 0.01% or more. The units of f are kHz. The max. range
within equations is 30 A or 2 A or 10 V.

Line filter

Select OFF, 500 Hz, 5.5 kHz, or 50 kHz.

Frequency filter

Select OFF, or ON

A/D converter

Simultaneous voltage and current conversion and 16-bit resolution.
Conversion speed (sampling rate): Approximately 5

µ

s. See

harmonic measurement items for harmonic display.

Range switching

Can be set for each input element.

Auto range functions

Increasing range value
• When the measured values of U and I exceed 110% of the range

rating

• When the peak value exceeds approximately 330% of the range

rating (or approximately 660% for crest factor 6)

Decreasing range value
• When the measured values of U and I fall to 30% or less of the

range rating, and Upk and Ipk are 300% or less of the lower
range value (or 600% for crest factor 6)

Display

Display

8.4-inch color TFT LCD monitor

Total number of pixels*

640 (horiz.) x 480 (vert.) dots

Waveform display resolution

501 (horiz.) x 432 (vert.) dots

Same as the data update rate.
Exceptions are listed below.
• The display update interval of numeric display (4, 8, and 16 items) is 250 ms when the

data update rate is 50 ms or 100 ms.

• The display update interval of numeric display (ALL, Single List, and Dual List) is 500 ms

when the data update rate is 50 ms to 250 ms.

• The display update rate of the trend display, bar graph display, and vector display is 1 s

when the data update rate is 50 ms to 500 ms.

• The display update interval of the waveform display is approximately 1 s when the data

update rate is 50 ms to 1 s. However, it may be longer depending on the trigger setting.

* Up to 0.02% of the pixels on the LCD may be defective.

Calculation Functions

Note1) The instrument’s apparent power (S), reactive power (Q), power factor (l), and phase

angle (Ø) are calculated using measured values of voltage, current, and active power.
(However, reactive power is calculated directly from sampled data when TYPE3 is
selected.) Therefore, when distorted waveforms are input, these values may be different
from those of other measuring instruments based on different measuring principals.

Note 2) The value of Q in the Q

Σ

calculation is calculated with a preceding minus sign (-) when

the current input leads the voltage input, and a plus sign when it lags the voltage input,
so the value of Q

Σ

may be negative.

Waveform Display (WAVE display)