Supports crest factor 6, Waveform peak rms value – Atec Yokogawa-WT3000 User Manual

Page 11

background image

11

waveform
peak
RMS value

Crest factor

=

waveform peak

(CF, peak factor)

RMS value

Crest factor (CF) =

{measuring range

ϫ

CF setting (3 or 6)}

measured value (RMS)

The crest factor is the ratio of the waveform peak value and the RMS value.

SUPPORTS Crest Factor 6

When checking the measurable crest factor of our power measuring instruments,
please refer to the following equation.

* However, the peak value of the measured signal must be less than or equal to the continuous maximum allowed input

* The crest factor on a power meter is specified by how many
times peak input value is allowed relative to rated input value.
Even if some measured signals exist whose crest factors are
larger than the specifications of the instrument (the crest
factor standard at the rated input), you can measure signals
having crest factors larger than the specifications by setting a

measurement range that is large relative to the measured
signal. For example, even if you set CF = 3, CF5 or higher
measurements are possible as long as the measured value
(RMS) is 60% or less than the measuring range. Also, for a
setting of CF = 3, measurements of CF = 300 are possible with
the minimum effective input (1% of measuring range).

There are limitations on some specifications and functions. See the individual product catalogs for details.

0.02% of reading + 0.04% of range

DC, 0.1 Hz to 1 MHz

1, 2, 3, 4

15/30/60/100/150/300/600/1000[V] (when crest factor is 3)

7.5/15/30/50/75/150/300/500[V] (when crest factor is 6)

50m/100m/200m/500m/1/2/5/10[V] (when crest factor is 3)

25m/50m/100m/250m/500m/1/2.5/5[V] (when crest factor is 6)

1% to 130%



(ASSP)

(user-defined function)



2 channels (up to 8 channels with option /FQ)

(/G6)(opt.)

Torque, rotating speed input (motor version)(opt.)

(/G6)(opt.)

(20 functions)

600,000

999,999

99,999

8.4-inch TFT color LCD

Numerical values, waveforms, trends, bar graphs, vectors

Approximately 200 kS/s

(/G6)(opt.)

(/G6)(opt.)

(/G6)(opt.)(10cycle/50Hz, 12cycle/60Hz)

(/FL)(opt.)

(/CC)(opt.)

(/DT)(opt.)

20 channels (/DA)(opt.)

approximately 30MB

GP-IB; RS-232 (/C2)(opt.); USB (/C12)

VGA output (/V1)(opt.); Ethernet (/C7)(opt.)

Commands in IEEE488.2 standard

50m/100m/250m/500m/1/2/5/10/20[S]

PC card interface; USB (/C5)(opt.)

Built-in printer (front side) (/B5)(opt.)

Voltage, current, active power, reactive power, apparent power, power factor, phase angle, peak voltage, peak current, crest factor

Range

Measurement

parameters

Display

resolution

Display

Measurement/

functions

Other

features

None (communication commands vary from product to product)

WT3000

0.1% of reading + 0.05% of range

DC, 0.1 Hz ~ 1 MHz

1, 2, 3, 4, 5, 6

1.5/3/6/10/15/30/60/100/150/300/600/1000[V] (when crest factor is 3)

750m/1.5/3/5/7/5/15/30/50/75/150/300/500[V] (when crest factor is 6)

50m/100m/250m/500m/1/2.5/5/10[V] (when crest factor is 3)

25m/50m/125m/250m/500m/1.25/2.5/5[V] (when crest factor is 6)

1% to 110%




(user-defined function)



3 ch (up to 12 channels with option /FQ)


Torque and rotational velocity input(opt.)

(20)

60,000

999,999

99,999

8.4-inch TFT color LCD (XGA)

Numerical values, waveforms, trends, bar graphs, vectors

Approximately 2 MS/s

(/G5)(opt.)

(/G6)(opt.)

(/DT)(opt.)

20 channels (/DA)(opt.)

Approximately 32MB

GPIB, USB, Ethernet,

RGBOutput(V1)

Commands in IEEE488.2 standard

50m/100m/250m/500m/1/2/5/10/20[S]

USB

Built-in printer (front side)(opt.)

WT1800

(opt.):Optional

Direct input

External sensor
input

Basic power accuracy (50/60 Hz)

Measurement power bandwidth

Input elements

Voltage range

Current range

Guaranteed accuracy range for voltage and current ranges

Main measurement parameters

Peak hold (instantaneous maximum value hold)

MAX hold

Voltage RMS/MEAN simultaneous measurement

RMS/MEAN/AC/DC simultaneous measurement

Mean active power

Active power amount (WP)

Apparent power amount (WS)

Reactive power amount (WQ)

Frequency

Efficiency

Phase angle between phases (fundamental wave)

Motor evaluation

FFT spectral analysis

User-defined functions

Voltage, current, power

Power amount, current amount

Frequency

Display

Display format

Sampling frequency

Harmonic measurement

Dual Harmonic Measurement

Harmonic measurement in normal measurement mode

IEC standards-compliant harmonic measurement

Flicker measurement

Cycle by cycle measurement

Compensation function

Delta calculation function

DA output

Synchronized operation

Storage (internal memory for storing data)

Interfaces

Communication command compatibility

Communication command standards

Data updating interval

Removable storage

Printer

Select from 10m/20m/50m/100m/200m/500m/1/2/5[A]

or 1/2/5/10/20/50[A] (when crest factor is 3)

5m/10m/25m/50m/100m/250m/500m/1/2.5[A]

or 0.5/1/2.5/5/10/25[A] (when crest factor is 6)

Select from 0.5/1/2/5/10/20/30[A] or

5m/10m/20m/50m/100m/200m/500m/1/2 [A] (when crest factor is 3)

Select from 0.25/0.5/1/2.5/5/10/15[A] or

2.5/5m/10m/25m/50m/100m/250m/500m/1 [A] (when crest factor is 6)

Comparison of Specifications and Functions in WT3000, Other WT Series Models