9

Displayed average noise level of the R&S

®

FSW8

RF attenuation = 0 dB, termination = 50 Ω, normalized to 1 Hz RBW, trace average,
average mode log, sample detector, +5 °C to +40 °C, RF preamplifier off

2 Hz ≤ f ≤ 100 Hz

< –110 dBm, typ. –120 dBm

100 Hz < f ≤ 1 kHz

< –120 dBm, typ. –130 dBm

1 kHz < f < 9 kHz

< –135 dBm, typ. –147 dBm

RF attenuation = 0 dB, termination = 50 Ω, log. scaling, normalized to 1 Hz RBW,
RBW = 1 kHz, VBW = 1 Hz, +5 °C to +40 °C, RF preamplifier off,
without R&S

®

FSW-B25 electronic attenuator option

9 kHz ≤ f ≤ 1 MHz

< –145 dBm, typ. –150 dBm

1 MHz < f ≤ 1 GHz

< –150 dBm, typ. –154 dBm

1 GHz < f < 3 GHz

3

< –152 dBm, typ. –156 dBm

1 GHz < f < 3 GHz

4

< –155 dBm, typ. –160 dBm

3 GHz ≤ f ≤ 8 GHz

< –152 dBm, typ. –156 dBm

add 1 dB to the above values if R&S

®

FSW-B25 option is installed

RF attenuation = 0 dB, termination = 50 Ω, log. scaling, normalized to 1 Hz RBW,
RBW = 1 kHz, VBW = 1 Hz, +5 °C to +40 °C, with R&S

®

FSW-B24 option,

RF preamplifier = 30 dB, without R&S

®

FSW-B25 electronic attenuator option

10 MHz < f ≤ 50 MHz

–154 dBm (nom.)

50 MHz < f ≤ 150 MHz

< –163 dBm, typ. –166 dBm

150 MHz < f ≤ 8 GHz

< –166 dBm, typ. –169 dBm

add 1 dB to the above values if R&S

®

FSW-B25 option is installed

Improvement with noise cancellation

for noise-like signals

13 dB (nom.)

Displayed average noise level of the R&S

®

FSW13, R&S

®

FSW26, without R&S

®

FSW-B24 option

RF attenuation = 0 dB, termination = 50 Ω, normalized to 1 Hz RBW, trace average,
average mode log, sample detector, +5 °C to +40 °C

2 Hz ≤ f ≤ 100 Hz

< –110 dBm, typ. –120 dBm

100 Hz < f ≤ 1 kHz

< –120 dBm, typ. –130 dBm

1 kHz < f < 9 kHz

< –135 dBm, typ. –147 dBm

RF attenuation = 0 dB, termination = 50 Ω, log. scaling, normalized to 1 Hz RBW,
RBW = 1 kHz, VBW = 1 Hz, +5 °C to +40 °C, f ≥ 8 GHz: YIG preselector on,
without R&S

®

FSW-B25 electronic attenuator option

9 kHz ≤ f ≤ 1 MHz

< –145 dBm, typ. –150 dBm

1 MHz < f ≤ 1 GHz

< –149 dBm, typ. –154 dBm

1 GHz < f < 3 GHz

3

< –151 dBm, typ. –156 dBm

1 GHz < f < 3 GHz

4

< –154 dBm, typ. –159 dBm

3 GHz ≤ f < 8 GHz

< –151 dBm, typ. –156 dBm

8 GHz ≤ f < 13.6 GHz

< –150 dBm, typ. –155 dBm

13.6 GHz ≤ f < 18 GHz

< –149 dBm, typ. –153 dBm

18 GHz ≤ f < 25 GHz

< –147 dBm, typ. –150 dBm

25 GHz ≤ f < 26.5 GHz

< –143 dBm, typ. –146 dBm

add 1 dB to the above values for frequencies < 8 GHz, 2 dB for frequencies ≥ 8 GHz,
if R&S

®

FSW-B25 option is installed

RF attenuation = 0 dB, termination = 50 Ω, log. scaling, normalized to 1 Hz RBW,
RBW = 1 kHz, VBW = 1 Hz, +5 °C to +40 °C, YIG preselector off,
without R&S

®

FSW-B25 electronic attenuator option

8 GHz ≤ f < 13.6 GHz

< –150 dBm, typ. –155 dBm

13.6 GHz ≤ f < 25 GHz

< –149 dBm, typ. –153 dBm

25 GHz ≤ f < 26.5 GHz

< –147 dBm, typ. –150 dBm

add 2 dB to the above values if R&S

®

FSW-B25 option is installed

Improvement with noise cancellation

for noise-like signals

13 dB (nom.)