Specifications, Frequency – Atec Rohde-Schwarz-FSW User Manual
Page 4
Version 05.01, December 2012
4
Rohde & Schwarz R&S
®
FSW Signal and Spectrum Analyzer
Specifications
Frequency
Frequency range
R&S
®
FSW8
DC coupled
2 Hz to 8 GHz
AC coupled
10 MHz to 8 GHz
R&S
®
FSW13
DC coupled
2 Hz to 13.6 GHz
AC coupled
10 MHz to 13.6 GHz
R&S
®
FSW26
DC coupled
2 Hz to 26.5 GHz
AC coupled
10 MHz to 26.5 GHz
R&S
®
FSW43
DC coupled
2 Hz to 43.5 GHz
AC coupled
10 MHz to 43.5 GHz
Frequency resolution
0.01
Hz
Reference frequency, internal
Accuracy
±(time since last adjustment × aging rate
+ temperature drift + calibration accuracy)
Aging per year
standard
±1 × 10
–7
with R&S
®
FSW-B4 OCXO precision
frequency reference option
±3 × 10
–8
Temperature drift (0 °C to +50 °C)
standard
±1 × 10
–7
with R&S
®
FSW-B4 OCXO precision
frequency reference option
±1 × 10
–9
Achievable initial calibration accuracy
standard
±1 × 10
–8
with R&S
®
FSW-B4 OCXO precision
frequency reference option
±5 × 10
–9
Frequency readout
Marker resolution
1 Hz
Uncertainty
±(marker frequency × reference accuracy
+ 10 % × resolution bandwidth +
½ (span/(sweep points – 1)) + 1 Hz)
Number of sweep (trace) points
default value
1001
range
101 to 32001
Marker tuning frequency step size
marker step size = sweep points
span/(sweep points – 1)
marker step size = standard
span/(default sweep points – 1)
Frequency counter resolution
0.001 Hz
Count accuracy
±(frequency × reference accuracy +
½ (last digit))
Display range for frequency axis
0 Hz, 10 Hz to max. frequency
Resolution
0.1
Hz
Max. span deviation
±0.1 %
Spectral purity
SSB phase noise
frequency = 1000 MHz, carrier offset
10 Hz, without R&S
®
FSW-B4 option
–80 dBc (1 Hz) (nom.)
10 Hz, with R&S
®
FSW-B4 option
–90 dBc (1 Hz) (nom.)
100 Hz
< –100 dBc (1 Hz)
1 kHz
< –125 dBc (1 Hz)
10 kHz
< –134 dBc (1 Hz)
100 kHz
< –136 dBc (1 Hz)
1 MHz
< –145 dBc (1 Hz)
10 MHz
–155 dBc (1 Hz) (nom.)
Residual FM
frequency = 1000 MHz, RBW = 1 kHz,
sweep time = 100 ms
< 0.1 Hz (nom.)