General data – Atec Rohde-Schwarz-FSW User Manual
Page 20
Version 05.01, December 2012
20
Rohde & Schwarz R&S
®
FSW Signal and Spectrum Analyzer
General data
Display
30.7 cm (12.1”) WXGA color touchscreen
Resolution
1280 × 800 pixel (WXGA resolution)
Pixel failure rate
< 1 × 10
–5
Data storage
Internal
standard
solid state disk ≥ 8 Gbyte
External
supports USB 2.0 compatible memory
devices
Temperature
Temperature
operating temperature range
+5 °C to +50 °C
permissible temperature range
0 °C to +55 °C
storage temperature range
–40 °C to +70 °C
Climatic loading
+40 °C at 90 % rel. humidity,
in line with EN 60068-2-30,
without condensation
Mechanical resistance
Vibration
sinusoidal
5 Hz to 55 Hz
displacement: 0.15 mm constant
amplitude (1.8 g at 55 Hz);
55 Hz to 150 Hz
acceleration: 0.5 g constant
in line with EN 60068-2-6
random
10 Hz to 300 Hz, acceleration 1.2 g
(RMS), in line with EN 60068-2-64
Shock
40 g shock spectrum,
in line with MIL-STD-810E method
no. 516.4, procedure I,
MIL-PRF-28800F, class 3
EMC
in line with EMC Directive 2004/108/EC
including:
IEC/EN 61326-1
16,
17
IEC/EN 61326-2-1
CISPR 11/EN 55011
16
IEC/EN 61000-3-2
IEC/EN 61000-3-3
Recommended calibration interval
1
year
Power supply
AC input voltage range
100 V to 240 V
AC supply frequency
50 Hz to 60 Hz / 400 Hz,
Max. input current
7.3 A (100 V) to 4.6 A (240 V)
Power consumption
R&S
®
FSW8
150 W without options,
250 W with all options (meas.)
R&S
®
FSW13, R&S
®
FSW26
175 W without options,
275 W with all options (meas.)
R&S
®
FSW43
200 W without options,
300 W with all options (meas.)
Safety
in line with IEC 61010-1, EN 61010-1,
UL 61010-1,
CAN/CSA-C22.2 No. 61010-1-04
Test mark
VDE-GS,
C
CSA
US
16
Emission limits for class B equipment.
17
Immunity test requirement for industrial environment (EN 61326 table 2).