beautypg.com

Atec Rohde-Schwarz-FSW User Manual

Page 16

background image

Version 05.01, December 2012

16

Rohde & Schwarz R&S

®

FSW Signal and Spectrum Analyzer

Signal analysis bandwidth ≤ 80 MHz
Amplitude flatness

(1.25 × signal analysis bandwidth) ≤
f

center

< 8 GHz

±0.3 dB (nom.)

f

center

≥ 8 GHz, YIG preselector off

±0.5 dB (nom.)

Deviation from linear phase

(1.25 × signal analysis bandwidth) ≤
f

center

< 8 GHz

±1° (nom.)

f

center

≥ 8 GHz, YIG preselector off

±2° (nom.)

Level display nonlinearity

see “Display nonlinearity”

Level measurement uncertainty

see “Total measurement uncertainty,
YIG preselector off”

Third-order intermodulation distortion

see “Third-order intercept point (TOI)”

ADC related spurious response

mixer level = –30 dBm

12

analysis bandwidth < 17 MHz

–100 dBc (nom.)

17 MHz ≤ analysis bandwidth < 80 MHz

–80 dBc (nom.)

Other spurious responses

see

“Spurious responses”

Signal analysis bandwidth 80 MHz to 160 MHz

13

Amplitude flatness

YIG preselector off for f ≥ 8 GHz
150 MHz ≤ f

center

< 4 GHz

±0.5 dB (nom.)

4 GHz ≤ f

center

< 8 GHz

±0.7 dB (nom.)

8 GHz ≤ f

center

< 26.5 GHz

±1 dB (nom.)

26.5 GHz ≤ f

center

≤ 43.5 GHz

±2 dB (nom.)

Deviation from linear phase

YIG preselector off for f ≥ 8 GHz
150 MHz ≤ f

center

< 4 GHz

±1° (nom.)

4 GHz ≤ f

center

< 8 GHz

±2° (nom.)

8 GHz ≤ f

center

< 26.5 GHz

±2.5° (nom.)

26.5 GHz ≤ f

center

≤ 43.5 GHz

±4° (nom.)

Level display nonlinearity

0 dB to –70 dB

< 0.15 dB (nom.)

Level measurement uncertainty

add 0.2 dB (nom.) to the values in
“Total measurement uncertainty,
YIG preselector off”

Third-order intermodulation distortion

150 MHz ≤ f

center

< 8 GHz:

two –20 dBm tones at input mixer
within analysis bandwidth

12

,

f

center

≥ 8 GHz:

two –25 dBm tones at input mixer
within analysis bandwidth

12

,

reference level = signal level + 6 dB

–70 dBc (nom.)

Residual spurious response

RF attenuation 0 dB, f

center

≥ 150 MHz

–90 dBm (nom.)

ADC related spurious response

single tone within analysis bandwidth,
mixer level = –10 dBm

12

,

reference level = signal level,
f

center

≥ 150 MHz

–70 dBc (nom.)

Other spurious responses

see

“Spurious responses”

12

Level of a tone at the input mixer (also abbreviated as “mixer level”) = signal level – RF attenuation + preamplifier gain.

13

R&S

®