Atec Rohde-Schwarz-FSW User Manual
Page 8
Version 05.01, December 2012
8
Rohde & Schwarz R&S
®
FSW Signal and Spectrum Analyzer
Second-harmonic intercept point (SHI)
R&S
®
FSW8, R&S
®
FSW13, R&S
®
FSW26, RF attenuation = 0 dB, level = –5 dBm,
YIG preselector on, RF preamplifier off
1 MHz < f
in
≤ 350 MHz
> 50 dBm, typ. 62 dBm
350 MHz < f
in
≤ 500 MHz
> 70 dBm, typ. 80 dBm
500 MHz < f
in
< 1.5 GHz
3
> 47 dBm, typ. 52 dBm
500 MHz < f
in
< 1.5 GHz
4
> 62 dBm, typ. 70 dBm
1.5 GHz ≤ f
in
≤ 4 GHz
> 62 dBm, typ. 70 dBm
4 GHz < f
in
≤ 13.5 GHz
65 dBm (nom.)
R&S
®
FSW43, RF attenuation = 0 dB, level = –5 dBm,
YIG preselector on, RF preamplifier off
1 MHz < f
in
≤ 500 MHz
> 45 dBm, typ. 55 dBm
500 MHz < f
in
< 1.5 GHz
3
> 47 dBm, typ. 56 dBm
500 MHz < f
in
< 1.5 GHz
4
> 52 dBm, typ. 60 dBm
1.5 GHz ≤ f
in
≤ 4 GHz
> 62 dBm, typ. 70 dBm
4 GHz < f
in
≤ 21.75 GHz
65 dBm (nom.)
R&S
®
FSW8, R&S
®
FSW13, R&S
®
FSW26, R&S
®
FSW43 with R&S
®
FSW-B24 option,
RF attenuation = 0 dB, level = –50 dBm, YIG preselector on, RF preamplifier on
50 MHz < f
in
≤ 21.75 GHz
10 dBm (nom.)
3
Without R&S
®
FSW-B13 highpass filter option or highpass off.
4
With R&S
®
FSW-B13 highpass filter option, highpass on.