beautypg.com

Atec Rohde-Schwarz-FSH3-6-18 User Manual

Page 12

background image

Data Sheet

R&S FSH3, R&S FSH6, R&S FSH18

1145.5973.12

TI.12

E-14

Directional Power Sensor R&S FSH-Z44

Zero offset

after zeroing

± 4 mW

Range of typical measure-
ment error with modulation

FM, PM, FSK, GMSK
AM (80 %)
2 CW carriers with identical
power

/4-DQPSK

EDGE
cdmaOne, DAB
3GPP W-CDMA, cdma2000
DVB-T

*) if standard is selected

on the R&S FSH

0 % of measured value (0 dB)

± 3 % of measured value (± 0.13 dB)
± 2 % of measured value (± 0.09 dB)

± 2 % of measured value (± 0.09 dB)

± 0.5 % of measured value (± 0.02 dB) *)

± 1 % of measured value (± 0.04 dB) *)
± 2 % of measured value (± 0.09 dB) *)
± 2 % of measured value (± 0.09 dB) *)

Temperature coefficient

200 MHz to 300 MHz
300 MHz to 4 GHz

0.40 %/K (0.017 dB/K)
0.25 %/K (0.011 dB/K)

Max. peak envelope power

Power measurement range

DAB, DVB-T, cdmaOne,
cdma2000, 3GPP W-CDMA
Other signals at
video bandwidth

4 kHz

200 kHz

4 MHz

4 W to 300 W

0.4 W to 300 W

1 W to 300 W
2 W to 300 W