beautypg.com

Atec Rohde-Schwarz-FSH3-6-18 User Manual

Page 11

background image

R&S FSH3, R&S FSH6, R&S FSH18

Data Sheet

1145.5973.12

TI.11

E-14

Directional Power Sensor R&S FSH-Z44

Frequency range

200 MHz to 4 GHz

Power measurement range

30 mW to 300 W

VSWR referenced to 50

200 MHz to 3 GHz
3 GHz to 4 GHz

< 1.07
< 1.12

Power-handling capacity

depending on temperature
and matching
(see diagram below)

120 W to 1000 W

Insertion loss

200 MHz to 1.5 GHz
1.5 GHz to 4 GHz

< 0.06 dB
< 0.09 dB

Directivity

200 MHz to 3 GHz
3 GHz to 4 GHz

> 30 dB
> 26 dB

Average power

Power measurement range

CW, FM, PM, FSK, GMSK
3GPP W-CDMA,
cdmaOne, cdma2000,
DAB, DVB-T
Other modulated signals

CF: ratio of peak envelope
power to average power

30 mW to 300 W
30 mW to 120 W

30 mW to 300 W / CF

Measurement uncertainty

200 MHz to 300 MHz
300 MHz to 4 GHz

sine signal,
18 °C to 28 °C, no zero
offset

4.0 % of measured value (0.17 dB)
3.2 % of measured value (0.14 dB)