beautypg.com

Trigger functions – Atec Rohde-Schwarz-FSH-Z1 User Manual

Page 6

background image

Version 08.00, June 2011

6

Rohde & Schwarz R&S

®

FSH4/FSH8 Spectrum Analyzer

Level measurement uncertainty
Absolute level uncertainty at 100 MHz

+20 °C to +30 °C

< 0.3 dB

9 kHz ≤ f < 100 kHz
(models .04/.14/.08/.18 only)

nominal < 1.5 dB

100 kHz ≤ f < 10 MHz

nominal < 1.5 dB

10 MHz ≤ f ≤ 3.6 GHz

< 1 dB

Frequency response (+20 °C to +30 °C)

3.6 GHz < f ≤ 8 GHz

< 1.5 dB

Attenuator uncertainty

< 0.3 dB

Uncertainty of reference level setting

nominal < 0.1 dB

Display nonlinearity

SNR > 16 dB, 0 dB to –50 dB,
logarithmic level display

< 0.2 dB

Bandwidth switching uncertainty

reference: RBW = 10 kHz

nominal < 0.1 dB

95 % confidence level, +20 °C to +30 °C,
SNR > 16 dB, 0 dB to –50 dB below reference level, RF attenuation auto

10 MHz ≤ f ≤ 3.6 GHz

< 1 dB, typ. 0.5 dB

Total measurement uncertainty

3.6 GHz < f ≤ 8 GHz

< 1.5 dB, typ. 1 dB

Trigger functions

Trigger
Trigger source

free run, video, external

low → high transition

2.4 V

External trigger level threshold

high → low transition

0.7 V

Gated trigger
Gate source

external

Gate delay

10 µs to 100 s, min. resolution 10 µs
(or 1 % of delay)

Gate length

10 µs to 100 s, min. resolution 10 µs
(or 1 % of gate length)