beautypg.com

Atec Rohde-Schwarz-FSH-Z1 User Manual

Page 19

background image

Version 08.00, June 2011

Rohde & Schwarz R&S

®

FSH4/FSH8 Spectrum Analyzer 19

Max. peak envelope power
Power measurement range

4 kHz

0.4 W to 300 W

200 kHz

1 W to 300 W

Video bandwidth

600 kHz

2 W to 300 W

Measurement uncertainty

same as for average power plus effect of
peak hold circuit

+18 °C to +28 °C

Error limits of peak hold circuit for burst signals

video bandwidth 4 kHz

±(3 % of measured value + 0.05 W)
starting from a burst width of 200 µs

video bandwidth 200 kHz

±(3 % of measured value + 0.20 W)
starting from a burst width of 4 µs

Duty cycle ≥ 0.1 and
repetition rate ≥ 100/s

video bandwidth 600 kHz

±(7 % of measured value + 0.40 W)
starting from a burst width of 2 µs

20/s ≤ repetition rate < 100/s

plus ±(1.6 % of measured value + 0.15 W)

0.001 ≤ duty cycle < 0.1

plus ±0.10 W

25 MHz to 40 MHz

0.50 %/K (0.022 dB/K)

Temperature coefficient

40 MHz to 1 GHz

0.35 %/K (0.015 dB/K)

Load matching
Matching measurement range

Return loss

0 dB to 23 dB

VSWR

>

1.15

Minimum forward power

specifications complied with ≥ 0.4 W

0.06 W

120 mm × 95 mm × 39 mm
(4.72 in × 3.74 in × 1.53 in)

Dimensions (W × H × D)

connecting cable

1.5 m (59 in)

Weight

0.65 kg (1.43 lb)

0

25

20

15

10

5

–4

6

4

2

–2

0

Return loss/dB

Error limits/dB

Error limits for matching measurements.

0.2

0.4

0.7

1

2

3

4

100

200

400

600

800

1000

0

Frequency/GHz

Forward power/W

AVG –10 °C to +35 °C

SWR ≤ 3

AVG +35 °C to +50 °C

SWR ≤ 3

PEAK (max. 10 ms), SWR ≤ 3

AVG –10 °C to +35 °C

SWR ≤ 1.5

Power-handling capacity.