beautypg.com

Accessories, Fsh-z1 and r&s, Fsh-z18 power sensors – Atec Rohde-Schwarz-FSH-Z1 User Manual

Page 18: Fsh-z14 directional power sensor

background image

Version 08.00, June 2011

18

Rohde & Schwarz R&S

®

FSH4/FSH8 Spectrum Analyzer

Accessories

R&S

®

FSH-Z1 and R&S

®

FSH-Z18 power sensors

R&S

®

FSH-Z1

10 MHz to 8 GHz

Frequency range

R&S

®

FSH-Z18

10 MHz to 18 GHz

10 MHz to 30 MHz

< 1.15

30 MHz to 2.4 GHz

< 1.13

2.4 GHz to 8 GHz

< 1.20

VSWR

8 GHz to 18 GHz

< 1.25

average power

400 mW (+26 dBm)

Maximum input power

peak power (< 10 µs, 1 % duty cycle)

1 W (+30 dBm)

Measurement range

200 pW to 200 mW
(–67 dBm to +23 dBm)

Signal weighting

average power

Effect of harmonics

< 0.5 % (0.02 dB)
at harmonic ratio of 20 dB

Effect of modulation

< 1.5 % (0.07 dB)
for continuous digital modulation

Absolute measurement uncertainty

sine signals, no zero offset
+15 °C to +35 °C

< 2.3 % (0.10 dB)

10 MHz to 8 GHz

0 °C to +50 °C

< 4.2 % (0.18 dB)

+15 °C to +35 °C

< 3.5 % (0.15 dB)

8 GHz to 18 GHz

0 °C to +50 °C

< 5.0 % (0.21 dB)

Zero offset after zeroing

< 110 pW

48 mm × 31 mm × 170 mm
(1.9 in × 1.22 in × 6.7 in)

Dimensions (W × H × D)

connecting cable

1.5 m (59 in)

Weight

< 0.3 kg (0.66 lb)

R&S

®

FSH-Z14 directional power sensor

Frequency range

25 MHz to 1 GHz

Power measurement range

30 mW to 300 W

VSWR referenced to 50 Ω

<

1.06

Power-handling capacity

depending on temperature and matching
(see diagram on page 19)

100 W to 1000 W

Insertion loss

< 0.06 dB

Directivity

> 30 dB

Average power
Power measurement range

CW, FM, PM, FSK, GMSK

CF: ratio of peak envelope

30 mW to 300 W

Modulated signals

power to average power

30 mW to 300 W/CF

Measurement uncertainty

25 MHz to 40 MHz

sine signal

4.0 % of measured value (0.17 dB)

40 MHz to 1 GHz

+18 °C to +28 °C, no zero offset

3.2 % of measured value (0.14 dB)

Zero offset

after zeroing

±4 mW

FM, PM, FSK, GMSK

0 % of measured value (0 dB)

AM (80 %)

±3 % of measured value (±0.13 dB)

two CW carriers with identical power

±2 % of measured value (±0.09 dB)

Range of typical measurement error
with modulation

EDGE, TETRA

±0.5 % of measured value (±0.02 dB)

3

25 MHz to 40 MHz

0.40 %/K (0.017 dB/K)

Temperature coefficient

40 MHz to 1 GHz

0.25 %/K (0.011 dB/K)

3

If standard is selected on the R&S

®

FSH4/R&S

®