Fsh-k47 1xev-do, Bts pilot channel and evm measurement application – Atec Rohde-Schwarz-FSH-Z1 User Manual
Page 14
![background image](/manuals/319455/14/background.png)
Version 08.00, June 2011
14
Rohde & Schwarz R&S
®
FSH4/FSH8 Spectrum Analyzer
Composite EVM
SNR > 30 dB
Measurement range
–40 dBm < P
RF channel
< 20 dBm
1.5 % < EVM < 25 %
1.5 % < EVM ≤ 10%
typ. 2.0 %
Measurement uncertainty
10 % < EVM < 25 %
typ. 2.5 %
Residual EVM
typ. 2.5 %
Rho
SNR > 30 dB
Measurement range
–40 dBm < P
RF channel
< 20 dBm
0.9 < Rho < 1
0.97 < Rho ≤ 1.0
typ. 0.005
Measurement uncertainty
0.90 < Rho ≤ 0.97
typ. 0.02
R&S
®
FSH-K47 1xEV-DO
®
BTS pilot channel and EVM measurement
application
The specifications below apply to the R&S
®
FSH4 and R&S
®
FSH8. They are based on the data sheet specifications of the R&S
®
FSH4
and R&S
®
FSH8, have not been checked separately and are not verified during instrument calibration. Measurement uncertainties are
indicated as 95 % confidence intervals. The specified level measurement errors do not take into account systematic errors due to
reduced signal-to-noise ratio (SNR).
Measurements R&S
®
FSH-K47
Spectrum overview
●
Result summary
●
RF channel power
●
Pilot Rho
●
Carrier frequency error
●
Traffic activity
●
Pilot EVM
●
PN timing (tau)
●
Peak to average
●
Pilot power
●
MAC power
●
Data power
●
All specifications are valid for RC3, one traffic channel, SNR > 30 dB, +15 °C to +35 °C.
Frequency range
15 MHz to 3.0 GHz
Carrier frequency uncertainty, nominal values
Lock range
±5 kHz
Measurement uncertainty
SNR > 30 dB,
Δf
ref
= uncertainty of reference frequency
< 100 Hz + Δf
ref
RF channel power
frequency > 15 MHz
preamplifier = OFF
–60 dBm < P
RF channel
< 20 dBm
Measurement range
preamplifier = ON
–75 dBm < P
RF channel
< 20 dBm
Measurement uncertainty
–75 dBm < P
RF channel
< 20 dBm,
ref. level adjusted to RF channel power
< 1 dB, typ. 0.5 dB
Pilot power
SNR > 30 dB
Measurement range
–40 dBm < P
RF channel
< 20 dBm
P
RF channel
– 20 dB < P
PICH
< P
RF channel
Measurement uncertainty
P
RF channel
– 20 dBm < P
CPICH
< P
RF channel
< 1 dB, typ. 0.5 dB
MAC power
SNR > 30 dB
Measurement range
–40 dBm < P
RF channel
< 20 dBm
P
RF channel
– 20 dB < P
SYNC
< P
RF channel
Measurement uncertainty
P
RF channel
– 20 dBm < P
SYNC
< P
RF channel
< 1 dB, typ. 0.5 dB
Data power
SNR > 30 dB
Measurement range
–40 dBm < P
RF channel
< 20 dBm
P
RF channel
– 20 dB < P
SYNC
< P
RF channel
Measurement uncertainty
P
RF channel
– 20 dBm < P
SYNC
< P
RF channel
< 1 dB, typ. 0.5 dB
Pilot EVM
SNR > 30 dB
Measurement range
–40 dBm < P
RF channel
< 20 dBm
1.5 % < EVM < 25 %
1.5 % < EVM ≤ 10%
typ. 2.0 %
Measurement uncertainty
10 % < EVM < 25 %
typ. 2.5 %
Residual EVM
typ. 2.5 %
Pilot Rho
SNR > 30 dB
Measurement range
–40 dBm < P
RF channel
< 20 dBm
0.9 < Rho < 1
0.97 < Rho ≤ 1.0
typ. 0.005
Measurement uncertainty
0.90 < Rho ≤ 0.97
typ. 0.02