Version 08.00, June 2011

Rohde & Schwarz R&S

®

FSH4/FSH8 Spectrum Analyzer 15

R&S

®

FSH-K50/R&S

®

FSH-K51 LTE FDD/TDD downlink pilot channel and EVM

measurement application

2

R&S

®

FSH-K50E/R&S

®

FSH-K51E LTE FDD/TDD downlink extended channel

and modulation measurement application

2

The specifications below apply to the R&S

®

FSH4 and R&S

®

FSH8. They are based on the data sheet specifications of the R&S

®

FSH4

and R&S

®

FSH8, have not been checked separately and are not verified during instrument calibration. Measurement uncertainties are

indicated as 95 % confidence intervals. The specified level measurement errors do not take into account systematic errors due to
reduced signal-to-noise ratio (S/N).

Measurements R&S

®

FSH-K50/R&S

®

FSH-K51 R&S

®

FSH-K50E/R&S

®

FSH-K51E

Spectrum overview

Result summary

RF channel power

Carrier frequency error

I/Q offset

Cell identity

Cyclic prefix

Reference signal power

PSYNC power

SSYNC power

PBCH power

PCFICH power

PDSCH power

Reference signal EVM

PSYNC EVM

SSYNC EVM

PBCH EVM

PCFICH EVM

PDSCH EVM

Constellation diagram

PSYNC –

SSYNC –

QPSK –

16QAM –

64QAM –

BTS scanner

Cell identity

PSYNC power

SSYNC power


All specifications are valid for SNR > 30 dB, +15 °C to +35 °C.

2

R&S

®

FSH-K50/R&S

®

FSH-K51/R&S

®

FSH-K50E/R&S

®

FSH-K51E options require instruments with serial number ≥ 105000.

Frequency range

15 MHz to 3.0 GHz

Supported channel bandwidths

1.4/3/5/10/15/20

MHz

Carrier frequency uncertainty
Lock range

±10 kHz

Measurement uncertainty

SNR > 30 dB,
Δf

ref

= uncertainty of reference frequency

< 10 Hz + Δf

ref

RF channel power


frequency > 15 MHz

preamplifier = OFF

–60 dBm < P

RF channel

< 20 dBm

Measurement range

preamplifier = ON

–75 dBm < P

RF channel

< 20 dBm

Measurement uncertainty

–75 dBm < P

RF channel

< 20 dBm,

ref. level adjusted to RF channel power

< 1 dB, typ. 0.5 dB

EVM

Measurement range

–50 dBm < P

RF channel

< 10 dBm, 860 MHz < frequency < 2.69 GHz,

E-UTRA test model 3.1, bandwidth 10 MHz, reference signal and PDSCH

Residual EVM

< 2.5 %, typ. 2.0 %