P L Z - 4 W S E R I E S

Sequence function

Model

PLZ164W

PLZ334W

PLZ1004W

PLZ164WA

PLZ664WA

Normal
sequence

Operation mode

CC, CR, CV, or CP

Maximum number of steps

256

Step execution time

1 ms – 999 h 59 min

Time resolution

1 ms (1 ms – 1 min)/100 ms (1 min – 1 h)/1 s (1 h – 10 h)/

10 s (10 h – 100 h)/1 min (100 h – 999 h 59 min)

Fast
sequence

Operation mode

CC or CR

Maximum number of steps

1024

Step execution time

25

μs – 100 ms

Time resolution

25

μs (25 μs – 100 μs)/100 μs (100 μs – 100 ms)

Others, Common specifications

Model

PLZ164W

PLZ334W

PLZ1004W

PLZ164WA

PLZ664WA

Elapsed time display

Measures the time from load on to load off. On/Off selectable.
Measures from 1 s up to 999 h 59 min 59 s

Auto load off timer

Automatically turns off the load after a specified time elapses.
Can be set in the range of 1 s to 999 h 59 min 59 s or off

Front panel
BNC connector

TRIG OUT

Trigger output: Approx. 4.5 V, pulse width: Approx. 2

μs, output impedance: Approx. 500 Ω

Outputs a pulse during sequence operation and switching operation.

I MON OUT

Current monitor output
1 V f.s (H or L range) and 0.1 V f.s (M range)

Communication
function

GPIB

IEEE std. 488.1-1978
SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, C0, E1
Supports the SCPI and IEEE std. 488.2-1992 command set
Sets panel functions except the power switch and reads measured values

RS-232C

D-SUB 9-pin connector (conforms to EIA-232-D)
Sets panel functions except the power switch and reads measured values
Supports the SCPI and IEEE std. 488.2-1992 command set
Baud rate: 2400, 4800, 9600, 19200 bps
Data length: 8-bit, Stop bit: 1, 2-bit, Parity bit: None, Flow control: Xon/Xoff

USB

Conforms to USB 2.0 Specifications and USBTMC-USB488 Device Class Specifications
Sets panel functions except the power switch and reads measured values
Communication speed 12 Mbps (Full speed)

General Specifications

Model

PLZ164W

PLZ334W

PLZ1004W

PLZ164WA

PLZ664WA

Input voltage range

100 VAC – 240 VAC

(90 VAC – 250 VAC)

Single phase, continuous

100 VAC – 120 VAC/200 VAC – 240 VAC

(90 VAC – 132 VAC/180 VAC – 250 VAC)

Single phase

Input frequency range

47 Hz – 63 Hz

Power consumption

80 VAmax

90 VAmax

160 VAmax

450 VAmax

1500 VAmax

Inrush current

45 A

80 A

Operating temperature range

0 °C – 40 °C

Operating humidity range

20 % – 85 % RH (without condensation)

Storage temperature range

–25 °C – 70 °C

Storage humidity range

90 % RH or less (without condensation)

Isolation voltage

±500 V

Insulation
resistance

Primary - input terminal

500 VDC, 30 M

Ω or more (ambient humidity of 70 % RH or less)

Primary - chassis

500 VDC, 30 M

Ω or more (ambient humidity of 70 % RH or less)

Withstand
voltage

Primary - input terminal

No abnormalities at 1500 VAC for 1 minute.

Primary - chassis

No abnormalities at 1500 VAC for 1 minute.

Dimensions (mm)

See outline drawing

Weight

Approx. 7 kg

Approx. 8 kg

Approx. 15 kg

Approx. 7.5 kg

Approx. 16 kg

Battery backup

Backs up setup information

Accessories

Power cord

× 1 pc. (with SVT3, 18AWG, 3-pin plug, cable length of 2.4 m), Load input

terminal cover

× 1 piece (2 lock plates provided), Set of screws for the load input terminal

× 2 sets (bolts, nuts, and spring washers), Setup guide × 1 piece(Japanese,English),
Quick Reference(Japanese:1 piece, Englis:1 piece),CD-ROM × 1 piece

*

3

Electromagnetic compatibility
(EMC)

*

1

Conforms to the requirements of the following directive and standard.
EMC Directive 89/336/EEC
EN61326:1997/A2:2001
Emissions: Class A
Immunity: Minimum immunity test requirements
EN61000-3-2:2000
EN61000-3-3:1995/A1:2001

Safety

*

1

,

*

2

Conforms to the requirements of the following directive and standard.
Low Voltage Directive 73/23/EEC
EN61010-1:2001
Class I
Pollution degree 2

[General Specifications]

*1

Only on models that have CE

marking on the panel.

Not applicable to custom order
models.

*2

This instrument is a Class I

equipment. Be sure to ground
the protective conductor terminal
of the instrument.

The safety of the instrument is
n o t g u a r a n t e e d u n l e s s t h e
instrument is grounded properly.

*3

C o n t a i n s A p p l i c a t i o n &

Samples,User

,

s manual,the

Communication Interface Manual

and KI-VISA.