beautypg.com

Phased array module specifications, Eddy current modules specifications – Atec Panametrics-Olympus-Omniscan-MX User Manual

Page 13

background image

www.olympus-ims.com

13

Phased Array Module Specifications

(Applies to OMNI-M-PA16128)

Overall dimensions

(W x H x D)

244 mm x 182 mm x 57 mm

(9.6 in. x 7.1 in. x 2.1 in.)

Weight

1.2 kg (2.6 lb)

Connectors

1 OmniScan connector for phased-array

probes

2 BNC connectors (1 pulser/receiver,

1 receiver for conventional UT) (BNC not

available on models 32:32 and 32:128)

Number of focal laws

256

Probe recognition

Automatic probe recognition and setup

Pulser/Receiver

Aperture

16 elements*

Number of elements

128 elements

Pulser

Voltage

80 V per element

Pulse width

Adjustable from 30 ns to 500 ns, resolution

of 2.5 ns

Fall time

Less than 10 ns

Pulse shape

Negative square wave

Output impedance

Less than 25 Ω

Receiver

Gain

0 dB to 74 dB, maximum input signal

1.32 Vp-p

Input impedance

75 Ω

System bandwidth

0.75 MHz to 18 MHz (–3 dB)

Beamforming

Scan type

Azimuthal and linear

Scan quantity

Up to 8

Active elements

16*

Elements

128

Delay range transmission

0 µs to 10 µs in 2.5-ns increments

Delay range reception

0 µs to 10 µs in 2.5-ns increments

Data acquisition

Digitizing frequency

100 MHz (10 bits)

Maximum pulsing rate

Up to 10 kHz (C-scan)

Acquisition depth

29 meters in steel (L-wave), 10 ms with com-

pression. 0.24 meter in steel (L-wave), 81.9 µs

without compression

Data processing

Number of data points

Up to 8000

Real-time averaging

2, 4, 8, 16

Rectifier

RF, full wave, halfwave +, halfwave –

Filtering

Low-pass (adjusted to probe frequency), digital

filtering (bandwidth, frequency range)

Video filtering

Smoothing (adjusted to probe frequency range)

Data storage

A-scan recording (TOFD)

6000 A-scans per second (512-point, 8-bit

A-scan)

C-scan type data recording

I, A, B, up to 10 kHz (amplitude or TOF)

Maximum file size

Limited to available internal flash memory:

180 MB (or 300 MB optional)

Data visualization

A-scan refresh rate

Real time: 60 Hz

Volume-corrected S-scan

Up to 40 Hz

Data synchronization

On internal clock

1 Hz to 10 kHz

On encoder

On 1 or 2 axes

Programmable time-corrected gain (TCG)

Number of points

16 (1 TCG curve per channel for focal laws)

Alarms

Number of alarms

3

Conditions

Any logical combination of gates

Analog outputs

2

* Models 16:16, 16:16M, 16:64M, 32:32, and 32:128 also available

Eddy Current Modules Specifications

Eddy Current Array

Eddy Current

Overall dimensions

(W x H x D)

244 mm x 182 mm x 57 mm

(9.6 in. x 7.1 in. x 2.1 in.)

Weight

1.2 kg (2.6 lb)

Connectors

1 OmniScan

®

connec-

tor for eddy current

array probes

N/A

1 19-pin Fischer eddy current probe connector
1 BNC connector

Number of channels

32 channels with inter-

nal multiplexer

64 channels with exter-

nal multiplexer

4 channels

Probe recognition

Automatic probe recognition and setup

Generator

Number of generators

1 (with internal electronic reference)

Maximum voltage

12 Vp-p into 10 Ω

Operating frequency

20 Hz to 6 MHz

Bandwidth

8 Hz to 5 kHz (in single coil). Inversely propor-

tional to the time-slot duration and set by the

instrument in multiplexed mode.

Receiver

Number of receivers

1 to 4

Maximum input signal

1 Vp-p

Gain

28 dB to 68 dB
Internal multiplexer

Number of generators

32 (4 simultaneously

on 8 time slots; up

to 64 with external

multiplexer)

N/A

Maximum voltage

12 Vp-p into 50 Ω

Number of receivers

4 differential receivers

(8 time slots each)

Maximum input signal

1 Vp-p

Data acquisition

Digitizing frequency

40 MHz

Acquisition rate

1 Hz to 15 kHz (in single coil). The rate can

be limited by the instrument’s processing

capabilities or by delays set by the multiplexed

excitation mode.

A/D resolution

16 bits

Data processing

Phase rotation

0° to 360° with increments of 0.1°

Filtering

FIR low-pass, FIR high-pass, FIR band-pass,

FIR band-stop (adjustable cutoff frequency),

median filter (variable from 2 points to 200

points), mean filter (variable from 2 points to

200 points)

Channel processing

Mixing

Data storage

Maximum file size

Limited to available internal flash memory:

180 MB (or 300 MB optional)

Data synchronization

On internal clock

1 Hz to 15 kHz (single coil)

External pace

Yes

On encoder

On 1 or 2 axes

Alarms

Number of alarms

3

Alarm zone shape

Pie, inverted pie, box, inverted box, and ring

Output type

Visual, audio, and TTL signals

Analog outputs

1 (X or Y)