beautypg.com

Omniscan specifications, Omniscan mx specifications, Ultrasound module specifications – Atec Panametrics-Olympus-Omniscan-MX User Manual

Page 12

background image

12

OmniScan MX Specifications

Overall dimensions

(W x H x D)

321 mm x 209 mm x 125 mm

(12.6 in. x 8.2 in. x 5.0 in.)

Weight

4.6 kg (10.1 lb), including module and

one battery

Data storage

Storage devices

CompactFlash card, most standard USB

storage devices, or through fast Ethernet,

internal 32-MB DiskOnChip

Data file size

160 MB

I/O ports

USB ports

3

Speaker out

Yes

Microphone input

Yes

Video output

Video out (SVGA)

Video input

Video input (NTSC/PAL)

Ethernet

10/100 Mbps

I/O lines

Encoder

2-axis encoder line (quadrature, up,

down, or clock/direction)

Digital input

4 digital TTL inputs, 5 V

Digital output

4 digital TTL outputs, 5 V, 10 mA

Acquisition on/off switch

Remote acquisition enable TTL, 5 V

Power output line

5 V, 500 mA power output line (short-

circuit protected)

Alarms

3 TTL, 5 V, 10 mA

Analog output

2 analog outputs (12 bits) ±5 V in 10 kΩ

Pace input

5 V TTL pace input

Display

Display size

21 cm (8.4 in.) (diagonal)

Resolution

800 pixels x 600 pixels

Number of colors

16 million

Type

TFT LCD

Power supply

Battery type

Smart Li-ion battery

Number of batteries

1 or 2 (battery chamber accommodates

two hot-swappable batteries)

Battery life

Minimum 6 hours with two batteries;

minimum 3 hours per battery under

normal operating conditions

DC-in voltage

15 V to 18 V (min. 50 W)

Environmental specifications

Operating temperature range

0 °C to 40 °C; 0 °C to 35 ºC with 32:128 PA

(32 ºF to 104 ºF; 32 ºF to 95 ºF with

32:128 PA)

Storage temperature range

–20 °C to 70 °C (–4 ºF to 158 ºF)

Relative humidity

0 % to 95 % noncondensing. No air

intake; splashproof design.

Ultrasound Module Specifications

Overall dimensions

(W x H x D)

244 mm x 182 mm x 57 mm

(9.6 in. x 7.1 in. x 2.1 in.)

Weight

1 kg (2.2 lb)

Connectors

LEMO 00 (2, 4, or 8)

Pulser

Number of pulsers

2, 4, or 8

Pulse output

50 V, 100 V, 200 V, 300 V ±10 %

(variable pulse width)

Pulse width

Adjustable from 30 ns to 1000 ns ±10 %,

resolution of 2.5 ns

Fall time

Less than 7 ns

Pulse shape

Negative square wave

Output impedance

Less than 7 Ω

Receiver

Number of receivers

2, 4, or 8

Receiver gain range

0 dB to 100 dB, by steps of 0.1 dB

Maximum input signal

20 Vp-p (screen at 128 %)

Minimum sensitivity

200 µVp-p (screen at 128 %)

Noise referred to input

160 µVp-p (26 µV RMS) (128 %)

Input impedance

50 Ω

Input filter

(100 % bandwidth)

Centered at 1 MHz (1.5 MHz),

centered at 2 MHz (2.25 MHz),

centered at 5 MHz (4 MHz),

centered at 10 MHz (12 MHz),

centered at 15 MHz,

centered at 20 MHz;

0.25 MHz to 2.5 MHz, 2 MHz to

25 MHz BB

System bandwidth

0.25 MHz to 32 MHz (–3 dB)

Rectifier

Both, positive, negative

Mode

PE (pulse-echo), PC (pitch-catch), TT

(through-transmission). In PC mode the

maximum number of pulsers equals the

number of channels divided by 2.

Smoothing

Digital

DAC

Number of points

16

DAC range

Up to 40 dB

Maximum gain slope

20 dB/µs

Data acquisition

A-scan acquisition rate

6000 A-scans/s (512-point A-scan)

Maximum pulsing rate

1 channel at 12 kHz (C-scan)

Data processing

Real-time averaging

2, 4, 8, or 16

Gates

Quantity

3: I (synchro), A and B (measure)

Synchronization

I, A, B referenced on main bang; A and

B referenced on gate I (post-synchroni-

zation)

Data storage

A-scan recording (TOFD)

6000 A-scans/s (512-point A-scan)

(3 MB/s transfer rate)

C-scan type data recording

12 000 (A1, A2, A3, T1, T2, T3) (3 gates)

12 kHz

(lower frequency for corrosion mapping)

Data visualization

Refresh rate

60 Hz

Data synchronization

On time

1 Hz to 12 kHz

On encoder

On 1 or 2 axes divided into 1 to 65 536

steps

Alarms

Number of alarms

3

Conditions

Any logical combination of gates

Signal

Amplitude or TOF of gate A or B

OmniScan Specifications