13
■
Measurement Indices (optional input module types)
• Each module has two input channels.
• Besides logic modules (16 channels), Model 8847 comes standard with 16 logic inputs integrated in the
device.
Trigger functions
Trigger mode
MEMORY
(high-speed recording)
: Auto, Single, Repeat
RECORDER
(real-time recording)
: Single, Repeat
Trigger sources
CH1 to CH16 (analog), Standard Logic 16ch + Logic Unit (Max. 3 units
48 channels), External (a rise of 2.5V or terminal short circuit), Timer,
Manual (either ON or OFF for each source),
Logical AND/OR of sources
Trigger types
• Level: Triggering occurs when preset voltage level is crossed
(upwards or downwards).
• Voltage drop: Triggering occurs when voltage drops below
peak voltage setting (for 50/60 Hz AC power lines only).
• Window: Triggering occurs when window defined by upper
and lower limit is entered or exited.
• Period: Rising edge or falling edge cycle of preset voltage
value is monitored and triggering occurs when defined cycle
range is exceeded.
• Glitch: Triggering occurs when pulse width from rising or
falling edge of preset voltage value is underrun.
• Event setting: Event count is performed for each source, and
triggering occurs when a preset count is exceeded.
• Logic: 1, 0, ×, pattern setting,
Level setting resolution 0.1% of full scale (full scale = 20 divisions)
Trigger filter
OFF, setting range 0.1 to 10.0 divisions
(MEMORY: high-speed
recording)
,
ON (10 ms fixed)/OFF
(RECORDER: real-time recording)
Trigger output
Open collector (5 voltage output, active Low)
at Level setting: pulse width (Sampling period × data number
after trigger)
at Pulse setting: pulse width (2ms)
Other functions
Trigger priority (OFF/ON), pre-trigger function for capturing
data from before/after trigger event (memory), level display
during trigger standby, start and stop trigger for recorder (real-
time recording)
FFT function
(function available from version 2.00 onward)
Analysis mode
Storage waveform, linear spectrum, RMS spectrum, Power
spectrum, Cross power spectrum,
Auto-correlation function, Histogram, Transfer function, Cross-
correlation function, Impulse response, Coherence function,
Octave analysis
Analysis channels 1 or 2 selected channels out of all analog channels
Frequency range 133 mHz to 8 MHz, External, (resolution 1/400, 1/800, 1/2000,
1/4000)
No.of sampling points 1000, 2000, 5000, 10000 points
Window functions Rectangular, Hanning, Exponential
Averaging
function
Time axis / frequency axis simple averaging, Exponential
averaging, Peak hold (frequency axis),
Averaging times: 2 times to 4,096 times
Other functions
• Display screen: Single, Dual, Nyquist
• Auto, Manual, Report printing
Measurement
target
With use input unit
Measurement range
Resolution
Voltage
ANALOG UNIT 8966
100 mV f.s. to 400 V f.s.
50 µV
HIGH RESOLUTION UNIT 8968
100 mV f.s. to 400 V f.s.
3.125 µV
DC/RMS UNIT 8972
100 mV f.s. to 400 V f.s.
50 µV
Current
* Using separate power supply
Use with the Clamp 9272-10
(20A), 9277, 3273-50, 3276
20 A f.s.
Note: 2 V f.s. voltage measurement
10 mA
Current
* Using separate power supply
Use with the Clamp 9272-10
(200A), 9278, 3274
200 A f.s.
Note: 2 V f.s. voltage measurement
0.1 A
Current
* Using separate power supply
Use with the Clamp 9279
500 A f.s.
Note: 2 V f.s. voltage measurement
0.25 A
Current
* Using separate power supply
Use with the Clamp 3275
500 A f.s.
Note: 2 V f.s. voltage measurement
0.25 A
RMS AC voltage
DC/RMS UNIT 8972
100 mV f.s. to 400 V f.s.
50 µV
Temperature
(thermocouple input)
TEMP UNIT 8967
200 °C f.s. to 2000 °C f.s.
Note: Upper and lower limit values depend
on the thermocouple
0.01 °C
Frequency, rpm
FREQ UNIT 8970
20 Hz f.s. to 100 kHz f.s.
2 (kr/min) f.s. to 2000 (kr/min) f.s.
0.1 Hz
10 (r/min)
Commercial power
supply frequency
FREQ UNIT 8970
40 Hz f.s. to 60 Hz f.s.
50 Hz f.s. to 70 Hz f.s.
0.1 Hz
Pulse count
FREQ UNIT 8970
40 k counts f.s. to 20 M counts f.s.
20 counts
Pulse duty ratio
FREQ UNIT 8970
100 % f.s.
0.5 %
Pulse width
FREQ UNIT 8970
0.01 s f.s. to 2 s f.s.
50 µs
Vibration/stress
STRAIN UNIT 8969
400
με f.s. to 20000 με f.s.
0.016
με
Relay contacts,
voltage on/off
LOGIC UNIT 8973
—
—
■
Maximum Recording Time for the Memory Function
• Operation cannot be guaranteed when the time axis is longer than one year.
• Setting recording length to an arbitrary value allows increasing the 200,000 division limit up to a
maximum of 320,000 divisions in 1 division units.
■
PC Software Specifications
Note: With use of the
8847 , Wv ver 1.26 or later
Wave Viewer (Wv) Software
(Application disk CD-R, bundled accessory)
Functions
• Simple display of waveform file
• Text conversion: convert binary data file to text format,
with selectable space or tab separators in addition to CSV,
and specifiable section, thinning available
• Display format settings: scroll functions, enlarge/reduce
display, display channel settings
• Others: voltage value trace function, jump to cursor/trigger
position function
Compatible PC
operating systems
Windows 95/98/Me, Windows NT 4.0 (SP3 or later), 2000,
XP
Maximum recording time
increases depending on
number of channels used
Analog 16 ch
+ internal Logic 16 ch
Analog 8 ch
+ internal Logic 16 ch
Analog 4 ch
+ internal Logic 16 ch
Analog 2 ch
+ internal Logic 16 ch
Logic 48 ch
+ internal Logic 16 ch
Time axis Samp.
period
20,000 div
50,000 div
100,000 div
200,000 div
100,000 div
5μs/div
50ns
100ms
250ms
500ms
1s
500ms
10μs/div
100ns
200ms
500ms
1s
2s
1s
20μs/div
200ns
400ms
1s
2s
4s
2s
50μs/div
500ns
1s
2.5s
5s
10s
5s
100μs/div
1μs
2s
5s
10s
20s
10s
200μs/div
2μs
4s
10s
20s
40s
20s
500μs/div
5μs
10s
25s
50s
1min 40s
50s
1ms/div
10μs
20s
50s
1min 40s
3min 20s
1min 40s
2ms/div
20μs
40s
1min 40s
3min 20s
6min 40s
3min 20s
5ms/div
50μs
1min 40s
4min 10s
8min 20s
16min 40s
8min 20s
10ms/div
100μs
3min 20s
8min 20s
16min 40s
33min 20s
16min 40s
20ms/div
200μs
6min 40s
16min 40s
33min 20s
1h 06min 40s
33min 20s
50ms/div
500μs
16min 40s
41min 40s
1h 23min 20s
2h 46min 40s
1h 23min 20s
100 ms/div
1 ms
33 min 20 s
1 h 23 min 20 s
2 h 46 min 40 s
5 h 33 min 20 s
2 h 46 min 40 s
200 ms/div
2 ms
1 h 06 min 40 s
2 h 46 min 40 s
5 h 33 min 20 s
11 h 06 min 40 s
5 h 33 min 20 s
500 ms/div
5 ms
2 h 46 min 40 s
6 h 56 min 40 s
13 h 53 min 20 s
1 d 03 h 46 min 40 s
13 h 53 min 20 s
1 s/div
10 ms
5 h 33 min 20 s
13 h 53 min 20 s
1 d 03 h 46 min 40 s
2 d 07 h 33 min 20 s
1 d 03 h 46 min 40 s
2 s/div
20 ms
11 h 06 min 40 s
1 d 03 h 46 min 40 s
2 d 07 h 33 min 20 s
4 d 15 h 06 min 40 s
2 d 07 h 33 min 20 s
5 s/div
50 ms
1 d 03 h 46 min 40 s
2 d 21 h 26 min 40 s
5 d 18 h 53 min 20 s
11 d 13 h 46 min 40 s
5 d 18 h 53 min 20 s
10 s/div
100 ms
2 d 07 h 33 min 20 s
5 d 18 h 53 min 20 s
11 d 13 h 46 min 40 s
23 d 03 h 33 min 20 s
11 d 13 h 46 min 40 s
30 s/div
300 ms
6 d 22 h 40 min 00 s
17 d 08 h 40 min 00 s
34 d 17 h 20 min 00 s
69 d 10 h 40 min 00 s
34 d 17 h 20 min 00 s
50 s/div
500 ms
11 d 13 h 46 min 40 s
28 d 22 h 26 min 40 s
57 d 20 h 53 min 20 s 115 d 17 h 46 min 40 s
57 d 20 h 53 min 20 s
100 s/div
1.0 s
23 d 03 h 33 min 20 s
57 d 20 h 53 min 20 s 115 d 17 h 46 min 40 s 231 d 11 h 33 min 20 s 115 d 17 h 46 min 40 s
1 min/div
600 ms
13 d 21 h 20 min 00 s
34 d 17 h 20 min 00 s
69 d 10 h 40 min 00 s 138 d 21 h 20 min 00 s
69 d 10 h 40 min 00 s
2 min/div
1.2 s
27 d 18 h 40 min 00 s
69 d 10 h 40 min 00 s 138 d 21 h 20 min 00 s 277 d 18 h 40 min 00 s 138 d 21 h 20 min 00 s
5 min/div
3.0 s
69 d 10 h 40 min 00 s 173 d 14 h 40 min 00 s 347 d 05 h 20 min 00 s 694 d 10 h 40 min 00 s 347 d 05 h 20 min 00 s